PAGE
14
/
53
Step 4. Measure/Adjustment optimal Zero balance of Scale (
display)
Make empty the scale part, and press
key.
Indicator check the current Zero balance and save the value and move next step.
Step 5. Input the Rate Output (mV/V) value of load cell
Input Max. Output Rate(mV/V) value of load cell with No. keys.
- Under this step, input Max. Output rate(mV) of load cell.
- If you installed a few pieces of load cells, the connection will be parallel, so the rated output of
a few load cells are as same as single load cell
’s rated output.
-
The Output rate is stated on “Calibration certificate” or “Label”
And press
key.
Step 6. End Calibration and Auto Reset
- Calculated Span value will be displayed and automatically reset and move the normal
weight indicating mode.
CAL _0_