5-3. Simulation Calibration Mode (Without Test Weight)
– Mode 2.
- This calibration Method will be useful to make calibration more than 10ton capacity setting.
- Guaranteed resolution will be 1/5,000 and if you need higher resolution, please make
calibration with Test weight.
Step 1. Enter to the
“SET-CAL” mode
Please turn on the Power + with pressing
display
Press
key
display.
Press
key to start “Simulation Calibration Mode”.
Remarks : Go to next step with save
key / Back to previous step
key Step 2.
Step 2.
Digit (Division)setting (
display)
Press
key and stop the optimal Digit value.
(Whenever pressing
key, digit value will be changed, like 01-02-05-10-20-50)
Press
key and save change and move to next step.
Step 3. Max. Capacity of Load cell
(
display)
Input Max. Capacity of Scale with No. keys.
- Under this step, input Total sum of each load cell
’s Max. Capacity. (Not weighing Scale)
- The Max. Capacity of load cell is stated on
“Test report”
or
“Label”
.
- If you installed 4 load cells, and each load cell
’s Max. Capacity is 500kg, then you have to input
2,000kg, as a Max. Capacity.
Input Capacity and press
key, and move to next step.
XXXXXX
CAPA
d
_ 1
CAL_ _2
_1
SET-CAL
TEST