17
Lubricating
Adapter
fig. a
fig. b
The lubricant doesn’t follow the product.
Attention:
The non-use of the lubricant would result
in the loss of the product warranty.
Attention
Handpieces not subjected to sterilization process should be lubricated at least twice a day.
Spray lubrication
To carry out the lubrication of your equipment, we recommend the use of the Lubs product.
Lubrication must be carried out by the posterior hole of the counter angle. Insert the adapter into
the hole as illustrated in fig. (a).
Firmly fit the oil applicator nozzle into the adapter hole (b) and apply the spray for 1 or 2 seconds,
by observing the lubrication into the instrument head.
Cover the piece with absorbent paper to remove the excess of oil.
Warning
Before lubricating the handpieces, remove the bit.
EQUIPMENT REUSE