Sensibility to predictable environmental conditions in regular use conditions

- The equipment has been planned not to be sensitive to interference such as magnetic fields, 

external electrical factors, electrostatic discharge, pressure or variance of pressure, provided that

the  equipment  is  installed,  maintained,  clean,  preserved,  transported  and  operated  as  per  this 

instruction for use.

Precautions and warnings “during the use” of equipment

- The equipment should only be operated by duly enabled and trained technicians (Dental Surgeons,

Capacitated Professionals).

- Before the first use and/or after long interruptions from work such as vacations, clean and disinfect

the equipment.

- If any maintenance should be required, only use services of the Alliage Authorized Technical 

Assistance.

- Owing to its delicate components, this handpiece must not be dropped or undergo impacts.

- Although this equipment has been planned in accordance with the standards of electromagnetic

compatibility, it can, in very extreme conditions, cause interference with other equipment. Do not

use this equipment together with other devices very sensitive to interference or with devices which

create high electromagnetic disturbance.

- Do not expose the plastic parts to contact with chemical substances, use in the routines of dental

treatment, such as: acids, mercury, acrylic liquids, amalgams, etc.

- The use of pneumatic or electric tools to perform any type of recording on handpieces, could lead

to damage to the surface treatment of the product, causing the loss of warranty.

Precautions and warnings “after” the use of equipment

- Always maintain the equipment clean for the next operation.

- Do not modify any part of the equipment. Do not disconnect the cable or other connections 

without need.

- After using the equipment, clean and disinfect all the parts which may be in contact with the patient.

PRECAUTIONS, RESTRICTIONS AND WARNINGS

Manufacturer shall not be responsible for:

- Use of the equipment differing from that for which it is intended.

- Damages caused to the equipment, the professional and/or the patient by the incorrect installation

and erroneous procedures of maintenance, differing from those described in these Instructions for

use which come with the equipment or by the incorrect operation of it.

- If the equipment has any abnormality, check if the problem is related to any item listed in the topic

of unforeseen events (failures, causes and solutions). If it is not possible to resolve the problem, 

contact your representative (Alliage).

Attention 

The Equipment maintains its condition of safety and efficacy, provided that it is maintained

(stored) as mentioned in this instruction of use. Thus, the equipment will not lose or alter

its physical and dimensional features.

Warning 

Do not use the equipment in conditions unsuitable for use.

Precautions in case of alteration in the functioning of equipment

Summary of Contents for COBRA CONTRA ANGLE INTRA 32 CA

Page 1: ......

Page 2: ...ribuitor Alliage S A Ind strias M dico Odontol gica C N P J 55 979 736 0001 45 Insc Est 582 002 897 114 Rod Abr o Assed Km 53 450m Cx Postal 782 CEP 14097 500 Ribeir o Preto S P Brasil Telefone 55 16...

Page 3: ...tation storage and operation 14 Sensibility to predictable environmental conditions in regular use conditions 14 Precautions and warnings during the use of equipment 14 Precautions and warnings after...

Page 4: ...roblems Please read it and keep this with you Indication of Use The Alliage handpieces are intended to remove caries remove restorations and perform tooth sectioning as an aid in extracting teeth and...

Page 5: ...Application specification Alliage Dental Handpieces are designed to perform most routine dental procedures in intraoral region treatments Operation principle The Contra angle handpiece operates conne...

Page 6: ...Fastening system Plate locks LT Latch Type Internal Spray Irrigation System Fastening system PB Push Button Attention The contents of this page are of an informative nature the equipment being able t...

Page 7: ...spray adapter 02 Adapter for lubrication 03 Lubricant Lubs Attention The use of any part accessory or material not speci ed or foreseen in these instructions for use is entirely the user s responsibil...

Page 8: ...e up to 135 C Coupling type Type 2 Length X2 Long type min 32mm Torque 0 350 1 000 N cm TECHNICAL SPECIFICATIONS General features Bur Specifications As per DIN EN ISO 1797 1 1992 and made of steel or...

Page 9: ...9 Weights Model Net Weight Gross Weight 32 CA 50g 115g 32 CA PB 49g 114g 32 CA PB SPRAY 49g 114g SL 30 50g 115g SL 30 PB 49g 114g TECHNICAL SPECIFICATIONS...

Page 10: ...Dimensions mm 32 CA SL 30 32 CA PB 32 CA PB SPRAY SL 30 PB Bur Bur Visibility angle 8 Projection with Type 4 test mandrel Visibility angle 9 Projection with Type 4 test mandrel TECHNICAL SPECIFICATIO...

Page 11: ...ial waste collection site at the end of its useful life Attention Instructions for use Serial number Model number Manufacturing date Manufacturer To indicate a reference to spraying liquid ABBREVIATIO...

Page 12: ...ing on the limiter fig b Upon touching the limiter with the bur turn the lever to the start direction to lock the bur fig c On PB Push Button system fit the bur into the guide and turn it until it loc...

Page 13: ...piece to the appropriate hose connect it with the activated spray Check for a leak in the connection between the handpiece and the coupler Transportation storage and operation This equipment must be t...

Page 14: ...ylic liquids amalgams etc The use of pneumatic or electric tools to perform any type of recording on handpieces could lead to damage to the surface treatment of the product causing the loss of warrant...

Page 15: ...c organisms that must be stored in a yellow bag properly labeled with the symbol of biological hazard or stored in a container resistant to perforation tight and even collection and incineration The p...

Page 16: ...t detergent foam without alcohol that has active components based on didecyldimethylammonium chloride Apply the disinfectant detergent foam on the surface or on a clean cloth and spread it on the surf...

Page 17: ...he lubrication of your equipment we recommend the use of the Lubs product Lubrication must be carried out by the posterior hole of the counter angle Insert the adapter into the hole as illustrated in...

Page 18: ...ction description Recommended frequency Operating noise Audibly check the equipment for any ab normalities Daily Coupling Bur Check that the Bur bit is properly fixed ac cording to fixing instructions...

Page 19: ...of the spare parts replacement To ensure the lifetime of your device only replace original spare parts from Alliage They have the assurance of the standards and technical specifications required by t...

Page 20: ......

Reviews: