Operation Manual
NORM-O-TEMP
®
, Model 111Z
Page 29 of 50
D.
Press the DECREMENT
▼
button during the five second interval after pressing the TEMP
SET button.
1.
A beeper sounds each time it is pressed but does not beep when held down
continuously.
2.
The set point display changes; the numbers move down the scale. The lowest set point
is 20°C/68°F. However, the operating range of the device is from 32°C/89.6°F to
42°C/107.6°F.
E.
Press the INCREMENT
▲
button during the five second interval after pressing the TEMP
SET button.
1.
A beeper sounds each time it is pressed, but does not beep when held down
continuously.
2.
The set point display changes; the numbers move up the scale. The highest setting is
42°C/107°F.
F.
To complete this test routine, press the Power switch to the “O” position.
1.
The control panel goes blank.
2.
The water flow indicator stops.
G.
Disconnect the power cord, loosely coil it and attach it to the nylon straps on the back of
the unit.
H.
Check the blanket/pad for leaks. Water leaks present a risk of infection to the patient
because the water is not sterile. If a leak is found, the blanket/pad should not be used.
I.
Check the couplings at the unit and at the blanket/pad for positive connection.
J.
To drain the water from the reusable blanket/pad simply shut off power to the unit and
allow the water to drain from the blanket/pad back into the NORM-O-TEMP
®
unit.
For reusable PLASTIPAD blanket/pad(s), disconnect the connecting hose from the unit.
Loosely coil the hose lengthwise in the center of the blanket/pad. Fold the blanket/pad
lengthwise into the center, 1/3 from the left side and 1/3 from the right side.
For reusable Gelli-Roll blankets/pads either lay them flat or roll the blanket/pad. Do not
fold the Gelli-Roll blankets/pads as folding may damage blanket/pad.
For single-use blanket/pad(s) follow the instructions packaged with the blanket/pad.
The NORM-O-TEMP
®
unit, connecting hose, and blankets/pads are now ready for patient use.
3-3.
Unit and Patient Related Precautions
This unit requires both distilled water and electricity to operate.
NOTE:
Please consult the beginning of the manual for a complete list of warnings and cautions
related to the NORM-O-TEMP
®
system.
Summary of Contents for Norm-o-Temp 111Z
Page 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Page 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Page 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Page 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...