5-1.
Maintenance of Water Reservoir
The dual compartment reservoir holds approximately 1½ gallons (5.7 liters) of distilled water that
remains in the unit between periods of use.
Once a month the water reservoir should be drained and replenished. The drain hose, included in
the packaging of the unit, is attached and the water is pumped out of the unit. When the LO
WATER alarm sounds, the remaining water is drained by gravity. Distilled/sterile-distilled water
(or distilled water bacteriostatic agent preparation) is then poured into the reservoir.
Quarterly, at a minimum, the Internal Cleaning / Dry Storage Procedure, listed below, should be
conducted.
NOTE:
DO NOT USE DE-IONIZED WATER OR ALCOHOL.
Chemical Cleaning Circulation Chart
UNIT
MODE
TEMPERATURE
DURATION
NORM-O-TEMP 111Z
Heating
38°C
5 Minutes
5-1.1.
Internal Cleaning using Bleach / Dry Storage Procedure
For circulating water units utilizing household bleach.
The following procedure cleans the fluid circuit in these products. The required
tools/supplies are appropriate drain hose(s) and/or hose connector(s), household bleach,
distilled water, U.S.P. Grade Propylene Glycol, and the appropriate AC electrical power
Procedure:
A.
Drain the water from the reservoir as described in Section 5-1.4.
B.
Turn unit off (“O” position).
C.
Fill the reservoir with the appropriate amount of household bleach per chart below .
D.
Fill the reservoir with warm distilled water.
E.
Turn the unit on and circulate per the Chemical Cleaning Circulation Chart (above).
F.
Drain the unit as instructed in Step A.
CAUTION
•
For safe handling and use of chemicals follow manufacturer guidelines.
•
Always drain the NORM-O-TEMP
®
system to a sanitary drain because
bacteria
may be
present in the unit’s water supply.
UNIT
BLEACH
WATER
NORM-O-TEMP
®
, 111Z
168 Milliliters
5.4 Liters
Summary of Contents for Norm-o-Temp 111Z
Page 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Page 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Page 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Page 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...