EZT-550 Controller Manual
31
5.5 Timer Setup
The timer function allows you to program your chamber to start and stop automatically at predetermined times.
The timer can be set to initiate a profile or run the chamber at steady state conditions. To access the timer,
select Timer Setup from the Setup menu.
1) Begin by setting the timer mode of operation. Press the Mode Setup button to access the timer mode
settings screen.
Timer Type
The timer type selection determines what function the timer is to perform.
Constant SP:
Operates the chamber at the current setpoint entered for temperature (humidity) prior
to turning on the timer.
Profile:
Starts the last profile that was run prior to turning on the timer.