background image

Ÿ

Automatic defrosting function: The machine automatically defrosts according to its 

own needs, no need to enter the prescribed time or any other settings.

Ÿ

Energy saving control: The advanced Polyurethene material foaming process 

improves the insulation performance, reduces energy consumption, and saves more 
electrical energy.

Ÿ

DC brushless motor drive: Low noise, high efficiency, no maintenance, long life, 

minimal heat, no effect on the chamber temperature.

Ÿ

Double cross flow fan driving air circulation: Wind pressure is high, large air volume, 

low noise, low heating capacity and long life.

Ÿ

12 automatic programmable sections control function: User can set 12 different 

speeds, temperatures, and times to automatically run the instrument at.

Ÿ

Acousto-optic alarm and protection functions: Over temperature alarm, over speed 

alarm, sensor fault alarm, timing alarm, open door alarm, over CO2 concentration 
alarm, over humidity alarm, electric leakage protection, no intake alarm (Instrument 
with CO2 function)

Ÿ

Open door protection: When the door opens, the machine stops running to protect the 

safety of the operator.

Ÿ

Automatic recovery run: When the operation stops as a result of power failure, after 

the power is supplied again, the instrument automatically resumes operation 
according to the original parameters of operation.

Ÿ

Long timing: User can set 0-999 hours and 59 minutes of operation time; the touch 

screen displays the remaining time, and when time runs up the instrument prompts 
"Program Finish".

Ÿ

Powerful data storage function: Data can be stored for more than 2 years. Data is 

recorded in 1 minute intervals.

Ÿ

Real-time data query function: On the display, operator can query the running data 

from the latest 7 days. Data is recorded in 1 minute intervals.

Ÿ

U disk export data function: With a USB port, it is easy to export more than 2 years of 

running data records to the USB.

 

Ÿ

Alarm information recording and query function: Alarm information memory is stored 

and can be queried.

Ÿ

Display has a real-time temperature curve display function: Historical data and real-

time data is displayed on the same diagram display and is easy to check.

Ÿ

Display can be switched from English interface to Chinese interface: If user is used to 

the Chinese interface, it is easy to switch at any time.

Ÿ

Remote WiFi monitor accessory: Instrument can be remotely controlled through 

wireless LAN monitoring or internet mobile phone. Many mobile phones can monitor 
the same instrument.

Ÿ

Computer monitoring software accessory: Instrument can be monitored by a LAN 

wired computer monitoring, and can remotely send alarm messages to connected 
mobile phones.

1. Overview

1. Overview

3

4

And more on the instrument version with CO2 functions:

Ÿ

CO2 infrared sensor: Accurate control, sensitive response, stable performance and 

long life. It is not affected by any changes in temperature or humidity. Match the full 
set of CO2 control accessories.

Ÿ

Multi High-Efficiancy-Particulate-Air (HEPA) filters: Prevent gas impurities in the 

cylinder or pollution source from contaminating the gas in the chamber.

Ÿ

Easy cleaning inner chamber: It is easy to wipe and disinfect the inner chamber 

without dismantling the platform to minimize the growth of bacteria.

Ÿ

Double hollow glass window: Heating and antifogging design allow the cell culture in 

the chamber to be clear at a glance.

Ÿ

High humidity, anti-rust, and anti-condensation design: Prevents the door from rusting 

due to high humidity. Prevents rust in the internal chamber. Prevents the micro flow of 
water on the door to the ground when the door is open.

Ÿ

Automatically liquid filled device accessory: The liquid can be automatically filled to 

the wet disk when the passive humidifying liquid is evaporated and dried in the 
chamber.

Ÿ

Active humidification device accessory: The humidity in the chamber can be 

controlled.

Procedure for testing the CO2 control function unit with 

RH Controller:

In order to ensure the stability of the humidity in the chamber, it is recommended that 
flasks be arranged evenly and that circulation and ventilation is ensured in the chamber.

The specific layout method and the number of flasks are as follows:

15 

12 

80%~85%

70%

 Set

RH Value

Max. QTY

of the Flask

Place

Test conditions

Take 250ml flask as an example:

1. Fill 250ml flask with 100ml pure water

2. Set the temperature of the instrument:

     

37

3. Set the speed of the instrument:  25rpm

1

4. 

Ambient temperature: 20

~25

Summary of Contents for AEOLUS IS-18A

Page 1: ...AEOLUS Stackable Incubator Shaker...

Page 2: ...odern biotechnology Our state of the art embedded microprocessor system utilized in this machine has powerful data processing ability outstanding stability excellent interference resistance and high p...

Page 3: ...odern biotechnology Our state of the art embedded microprocessor system utilized in this machine has powerful data processing ability outstanding stability excellent interference resistance and high p...

Page 4: ...antages of lightness sound insulation shockproof electrical insulation heat resistance cold resistance solvent resistance and aging resistance Automatic adjustment of PID control Instrument adjusts it...

Page 5: ...antages of lightness sound insulation shockproof electrical insulation heat resistance cold resistance solvent resistance and aging resistance Automatic adjustment of PID control Instrument adjusts it...

Page 6: ...am display and is easy to check Display can be switched from English interface to Chinese interface If user is used to the Chinese interface it is easy to switch at any time Remote WiFi monitor access...

Page 7: ...am display and is easy to check Display can be switched from English interface to Chinese interface If user is used to the Chinese interface it is easy to switch at any time Remote WiFi monitor access...

Page 8: ...x Humidity Door Open Direction Programmable Segments 12 Segments Relative Humidity Type Active Humidification Disinfection UV Lamps on Left and Right Walls Parameter Protection Screen Data Display Pas...

Page 9: ...x Humidity Door Open Direction Programmable Segments 12 Segments Relative Humidity Type Active Humidification Disinfection UV Lamps on Left and Right Walls Parameter Protection Screen Data Display Pas...

Page 10: ...nly Power Cord Liquid Collection Tray Fuse Cross Screwdriver 2 Single open end Wrench 24 Wrench S10 User Manual Filter Filter Protection Package Hood PU Pipe 5 m Silicone Tube 12 2 T Connectors 14 Two...

Page 11: ...nly Power Cord Liquid Collection Tray Fuse Cross Screwdriver 2 Single open end Wrench 24 Wrench S10 User Manual Filter Filter Protection Package Hood PU Pipe 5 m Silicone Tube 12 2 T Connectors 14 Two...

Page 12: ...The instrument must be placed on a firm and level surface The total weight of the instrument when fully loaded 474 lb unit must be taken into consideration There must be sufficient space left around t...

Page 13: ...The instrument must be placed on a firm and level surface The total weight of the instrument when fully loaded 474 lb unit must be taken into consideration There must be sufficient space left around t...

Page 14: ...eparation 1060mm 530mm 1290mm 880mm 340mm 530mm 530mm 1400mm 1290mm 530mm 530mm 530mm 1332mm 800mm 270mm 2055mm 80 mm Note When double and triple stacked it is recommended that the second and third in...

Page 15: ...eparation 1060mm 530mm 1290mm 880mm 340mm 530mm 530mm 1400mm 1290mm 530mm 530mm 530mm 1332mm 800mm 270mm 2055mm 80 mm Note When double and triple stacked it is recommended that the second and third in...

Page 16: ...he power box Changing the fuse Malfunctioning of the instrument Instrument is out of use for a long period of time 4 Safety Instructions 5 Installation Lifting the instrument by hand should be prohibi...

Page 17: ...he power box Changing the fuse Malfunctioning of the instrument Instrument is out of use for a long period of time 4 Safety Instructions 5 Installation Lifting the instrument by hand should be prohibi...

Page 18: ...filter first and then to the CO2 inlet located on the right side of the shaker See Figure 1 3 Set the CO2 secondary flow control valve to 0 08 0 1MPa 11 6 14 5psi WARNING Excessive pressure may cause...

Page 19: ...filter first and then to the CO2 inlet located on the right side of the shaker See Figure 1 3 Set the CO2 secondary flow control valve to 0 08 0 1MPa 11 6 14 5psi WARNING Excessive pressure may cause...

Page 20: ...d access port and tube and filter Note 5 Installation 17 18 Polyurethane Pipe Inner dia 7mm outer dia 10mm To the filter Stainless steel hoop To the CO2 regulator 6 Structure Down swing door Up swing...

Page 21: ...d access port and tube and filter Note 5 Installation 17 18 Polyurethane Pipe Inner dia 7mm outer dia 10mm To the filter Stainless steel hoop To the CO2 regulator 6 Structure Down swing door Up swing...

Page 22: ...e password needs to be input before some functions are modified The default password is 88888888 1 2 3 4 4 5 6 7 8 9 10 11 Main Interface 1 Displays the current status of the instrument 2 Light button...

Page 23: ...e password needs to be input before some functions are modified The default password is 88888888 1 2 3 4 4 5 6 7 8 9 10 11 Main Interface 1 Displays the current status of the instrument 2 Light button...

Page 24: ...3 Click the corresponding button to input the desired parameter value or click the corresponding ON OFF to control the instrument related function switch 4 After setting press Save and back or press...

Page 25: ...3 Click the corresponding button to input the desired parameter value or click the corresponding ON OFF to control the instrument related function switch 4 After setting press Save and back or press...

Page 26: ...address is valid when multi instrument networking 4 Pause when door is open Set the status to ON to keep the instrument running when the door is opened set the status to OFF to pause the instrument w...

Page 27: ...address is valid when multi instrument networking 4 Pause when door is open Set the status to ON to keep the instrument running when the door is opened set the status to OFF to pause the instrument w...

Page 28: ...dity inside chamber 4 Input the difference between set humidity and actual humidity with or 5 Repeat steps 1 4 for 50 70 and 90 0 0 0 0 0 0 0 0 1 0 3 0 5 0 10 0 c CO2 Calibrate CO2 Concentration can b...

Page 29: ...dity inside chamber 4 Input the difference between set humidity and actual humidity with or 5 Repeat steps 1 4 for 50 70 and 90 0 0 0 0 0 0 0 0 1 0 3 0 5 0 10 0 c CO2 Calibrate CO2 Concentration can b...

Page 30: ...CO2 The No Gas Alarm activates when the instrument does not receive a gas inject CO2 for over 5 minutes The main interface will show alarm information When an abnormality occurs in the instrument plea...

Page 31: ...CO2 The No Gas Alarm activates when the instrument does not receive a gas inject CO2 for over 5 minutes The main interface will show alarm information When an abnormality occurs in the instrument plea...

Page 32: ...ar Alarm History Clear all the historical alarm records saved by the instrument automatically 3 View Past Week s Data View the operation data of the instrument within the past week on the touchscreen...

Page 33: ...ar Alarm History Clear all the historical alarm records saved by the instrument automatically 3 View Past Week s Data View the operation data of the instrument within the past week on the touchscreen...

Page 34: ...ing under high humidity 18 Please make sure the air circulation is sufficient when using CO2 The CO2 will increase in a small confined room and can harm the operator s health Also please avoid direct...

Page 35: ...ing under high humidity 18 Please make sure the air circulation is sufficient when using CO2 The CO2 will increase in a small confined room and can harm the operator s health Also please avoid direct...

Page 36: ...is activated the instrument will continue to attempt to inject CO2 gas when the gas supply is resumed the alarm will be automatically cancelled Instructions Touchscreen off and Platform not shaking P...

Page 37: ...is activated the instrument will continue to attempt to inject CO2 gas when the gas supply is resumed the alarm will be automatically cancelled Instructions Touchscreen off and Platform not shaking P...

Page 38: ...om of the monitor iOS Android Scan QR code to download 12 2 Locking and Unlocking the Universal Aluminum Platform 1 When the instrument is not running unscrew the handle counterclockwise as in step 1...

Page 39: ...om of the monitor iOS Android Scan QR code to download 12 2 Locking and Unlocking the Universal Aluminum Platform 1 When the instrument is not running unscrew the handle counterclockwise as in step 1...

Page 40: ...s 13 Optional Accessories Model Name Specifications IS A50 IS A1 IS A2 IS A3 IS A4 IS A5 IS A6 IS A7 IS A8 IS A9 25ml 50ml 100ml 150ml 200ml 250ml 500ml 1000ml 2000ml 3000ml IS A11 IS A12 IS A23 IS A2...

Page 41: ...s 13 Optional Accessories Model Name Specifications IS A50 IS A1 IS A2 IS A3 IS A4 IS A5 IS A6 IS A7 IS A8 IS A9 25ml 50ml 100ml 150ml 200ml 250ml 500ml 1000ml 2000ml 3000ml IS A11 IS A12 IS A23 IS A2...

Page 42: ...antlement 4 No valid invoice except the goods can prove the validity of the warranty 5 Product model and serial number of products does not match the warranty certificate 6 The damage caused by force...

Page 43: ...AEOLUS Stackable Incubator Shaker...

Reviews: