Pay attention to avoid any risk of body injuries from disassembling, lifting, or 

moving the instrument.

Ÿ

The CO2 concentration may increase and can be harmful to your health if 

operating the instrument in a small and airtight room. 

Ÿ

Please keep the room ventilated when the CO2 function is on. Additionally, please 

avoid direct inhalation of the gas when opening the door.

Ÿ

Keep the vent of the instrument away from heat sources or other vents, otherwise 

the compressor cannot be started or started frequently during refrigeration.

Ÿ

When the instruments are stacked, make sure the floor loading capacity is at 

least 1.5 times of the instruments weight. Stack them up near the wall or on the 
floor with a reinforced beam to ensure good loading capacity.

The instrument must be placed on a firm and level surface. The total weight of the 

instrument when fully loaded (474 lb/unit) must be taken into consideration. There 

must be sufficient space left around the instrument. There should be at least 20 cm 

between the instrument and any wall or any two adjacent instruments.

3.1 Placement Conditions

Warning

3.2 Ambient Conditions

a. F

 

or indoor use only

b. A

 

mbient Temperature: 10

-35

c. R

 

elative Humidity: 20%-80% RH

d

.

Atmospheric Pressure: 75kPa-106kPa

e. Use a dedicated power outlet with a capacity no less than 1.3 kW and AC 110 V 

f. The fluctuations of power supply voltage is less than 10% of rated voltage

3. Preparation

3. Preparation

3.3 Space Conditions

10

3 Down-swing-door instruments are stacked as below:

1290mm

270mm

800mm

880mm

80

530mm

530mm

530mm

1670mm

mm

1330mm

Note: 

When double and triple stacked, it is recommended that the second and third 

instruments are not used at speeds exceeding 250 RPM.

Summary of Contents for AEOLUS IS-18A

Page 1: ...AEOLUS Stackable Incubator Shaker...

Page 2: ...odern biotechnology Our state of the art embedded microprocessor system utilized in this machine has powerful data processing ability outstanding stability excellent interference resistance and high p...

Page 3: ...odern biotechnology Our state of the art embedded microprocessor system utilized in this machine has powerful data processing ability outstanding stability excellent interference resistance and high p...

Page 4: ...antages of lightness sound insulation shockproof electrical insulation heat resistance cold resistance solvent resistance and aging resistance Automatic adjustment of PID control Instrument adjusts it...

Page 5: ...antages of lightness sound insulation shockproof electrical insulation heat resistance cold resistance solvent resistance and aging resistance Automatic adjustment of PID control Instrument adjusts it...

Page 6: ...am display and is easy to check Display can be switched from English interface to Chinese interface If user is used to the Chinese interface it is easy to switch at any time Remote WiFi monitor access...

Page 7: ...am display and is easy to check Display can be switched from English interface to Chinese interface If user is used to the Chinese interface it is easy to switch at any time Remote WiFi monitor access...

Page 8: ...x Humidity Door Open Direction Programmable Segments 12 Segments Relative Humidity Type Active Humidification Disinfection UV Lamps on Left and Right Walls Parameter Protection Screen Data Display Pas...

Page 9: ...x Humidity Door Open Direction Programmable Segments 12 Segments Relative Humidity Type Active Humidification Disinfection UV Lamps on Left and Right Walls Parameter Protection Screen Data Display Pas...

Page 10: ...nly Power Cord Liquid Collection Tray Fuse Cross Screwdriver 2 Single open end Wrench 24 Wrench S10 User Manual Filter Filter Protection Package Hood PU Pipe 5 m Silicone Tube 12 2 T Connectors 14 Two...

Page 11: ...nly Power Cord Liquid Collection Tray Fuse Cross Screwdriver 2 Single open end Wrench 24 Wrench S10 User Manual Filter Filter Protection Package Hood PU Pipe 5 m Silicone Tube 12 2 T Connectors 14 Two...

Page 12: ...The instrument must be placed on a firm and level surface The total weight of the instrument when fully loaded 474 lb unit must be taken into consideration There must be sufficient space left around t...

Page 13: ...The instrument must be placed on a firm and level surface The total weight of the instrument when fully loaded 474 lb unit must be taken into consideration There must be sufficient space left around t...

Page 14: ...eparation 1060mm 530mm 1290mm 880mm 340mm 530mm 530mm 1400mm 1290mm 530mm 530mm 530mm 1332mm 800mm 270mm 2055mm 80 mm Note When double and triple stacked it is recommended that the second and third in...

Page 15: ...eparation 1060mm 530mm 1290mm 880mm 340mm 530mm 530mm 1400mm 1290mm 530mm 530mm 530mm 1332mm 800mm 270mm 2055mm 80 mm Note When double and triple stacked it is recommended that the second and third in...

Page 16: ...he power box Changing the fuse Malfunctioning of the instrument Instrument is out of use for a long period of time 4 Safety Instructions 5 Installation Lifting the instrument by hand should be prohibi...

Page 17: ...he power box Changing the fuse Malfunctioning of the instrument Instrument is out of use for a long period of time 4 Safety Instructions 5 Installation Lifting the instrument by hand should be prohibi...

Page 18: ...filter first and then to the CO2 inlet located on the right side of the shaker See Figure 1 3 Set the CO2 secondary flow control valve to 0 08 0 1MPa 11 6 14 5psi WARNING Excessive pressure may cause...

Page 19: ...filter first and then to the CO2 inlet located on the right side of the shaker See Figure 1 3 Set the CO2 secondary flow control valve to 0 08 0 1MPa 11 6 14 5psi WARNING Excessive pressure may cause...

Page 20: ...d access port and tube and filter Note 5 Installation 17 18 Polyurethane Pipe Inner dia 7mm outer dia 10mm To the filter Stainless steel hoop To the CO2 regulator 6 Structure Down swing door Up swing...

Page 21: ...d access port and tube and filter Note 5 Installation 17 18 Polyurethane Pipe Inner dia 7mm outer dia 10mm To the filter Stainless steel hoop To the CO2 regulator 6 Structure Down swing door Up swing...

Page 22: ...e password needs to be input before some functions are modified The default password is 88888888 1 2 3 4 4 5 6 7 8 9 10 11 Main Interface 1 Displays the current status of the instrument 2 Light button...

Page 23: ...e password needs to be input before some functions are modified The default password is 88888888 1 2 3 4 4 5 6 7 8 9 10 11 Main Interface 1 Displays the current status of the instrument 2 Light button...

Page 24: ...3 Click the corresponding button to input the desired parameter value or click the corresponding ON OFF to control the instrument related function switch 4 After setting press Save and back or press...

Page 25: ...3 Click the corresponding button to input the desired parameter value or click the corresponding ON OFF to control the instrument related function switch 4 After setting press Save and back or press...

Page 26: ...address is valid when multi instrument networking 4 Pause when door is open Set the status to ON to keep the instrument running when the door is opened set the status to OFF to pause the instrument w...

Page 27: ...address is valid when multi instrument networking 4 Pause when door is open Set the status to ON to keep the instrument running when the door is opened set the status to OFF to pause the instrument w...

Page 28: ...dity inside chamber 4 Input the difference between set humidity and actual humidity with or 5 Repeat steps 1 4 for 50 70 and 90 0 0 0 0 0 0 0 0 1 0 3 0 5 0 10 0 c CO2 Calibrate CO2 Concentration can b...

Page 29: ...dity inside chamber 4 Input the difference between set humidity and actual humidity with or 5 Repeat steps 1 4 for 50 70 and 90 0 0 0 0 0 0 0 0 1 0 3 0 5 0 10 0 c CO2 Calibrate CO2 Concentration can b...

Page 30: ...CO2 The No Gas Alarm activates when the instrument does not receive a gas inject CO2 for over 5 minutes The main interface will show alarm information When an abnormality occurs in the instrument plea...

Page 31: ...CO2 The No Gas Alarm activates when the instrument does not receive a gas inject CO2 for over 5 minutes The main interface will show alarm information When an abnormality occurs in the instrument plea...

Page 32: ...ar Alarm History Clear all the historical alarm records saved by the instrument automatically 3 View Past Week s Data View the operation data of the instrument within the past week on the touchscreen...

Page 33: ...ar Alarm History Clear all the historical alarm records saved by the instrument automatically 3 View Past Week s Data View the operation data of the instrument within the past week on the touchscreen...

Page 34: ...ing under high humidity 18 Please make sure the air circulation is sufficient when using CO2 The CO2 will increase in a small confined room and can harm the operator s health Also please avoid direct...

Page 35: ...ing under high humidity 18 Please make sure the air circulation is sufficient when using CO2 The CO2 will increase in a small confined room and can harm the operator s health Also please avoid direct...

Page 36: ...is activated the instrument will continue to attempt to inject CO2 gas when the gas supply is resumed the alarm will be automatically cancelled Instructions Touchscreen off and Platform not shaking P...

Page 37: ...is activated the instrument will continue to attempt to inject CO2 gas when the gas supply is resumed the alarm will be automatically cancelled Instructions Touchscreen off and Platform not shaking P...

Page 38: ...om of the monitor iOS Android Scan QR code to download 12 2 Locking and Unlocking the Universal Aluminum Platform 1 When the instrument is not running unscrew the handle counterclockwise as in step 1...

Page 39: ...om of the monitor iOS Android Scan QR code to download 12 2 Locking and Unlocking the Universal Aluminum Platform 1 When the instrument is not running unscrew the handle counterclockwise as in step 1...

Page 40: ...s 13 Optional Accessories Model Name Specifications IS A50 IS A1 IS A2 IS A3 IS A4 IS A5 IS A6 IS A7 IS A8 IS A9 25ml 50ml 100ml 150ml 200ml 250ml 500ml 1000ml 2000ml 3000ml IS A11 IS A12 IS A23 IS A2...

Page 41: ...s 13 Optional Accessories Model Name Specifications IS A50 IS A1 IS A2 IS A3 IS A4 IS A5 IS A6 IS A7 IS A8 IS A9 25ml 50ml 100ml 150ml 200ml 250ml 500ml 1000ml 2000ml 3000ml IS A11 IS A12 IS A23 IS A2...

Page 42: ...antlement 4 No valid invoice except the goods can prove the validity of the warranty 5 Product model and serial number of products does not match the warranty certificate 6 The damage caused by force...

Page 43: ...AEOLUS Stackable Incubator Shaker...

Reviews: