Operation
17
English
3.3 Start up
When the
Xsafe Bright
is powered up, the unit will perform internal
diagnostic checks whilst the display will show a Crowcon logo. This procedure
will be displayed for about 45 seconds followed by displaying a warming up
status for about 120 seconds.
Diagram 5: Start up and Warming up
If the diagnostic checks were successful, the gas status screen will be
displayed. In normal operation the gas level will be indicated on the display.
Use the supplied magnetic wand to select the Scroll key to move between the
information screens shown below.
Diagram 6: Gas status screen
%
lel
50
ppm
16
%
vol
20.9
ma
4.00
oxygen
19.6
1419
20%vol
20.9
-0.0161
Gas name
PCB temperature
ADC value
Range
Gas value
Gain
SN: 0000000000000000
f: 2. 10 : 0.3 h: 151003
1 0 0 0
Volt: 23v dt: 20171206
1. Serial number
2. Firmware version, Hardware version
3. HART ID
4. Detector 485 Address, 485 packets
received, 485 packets address match,
Polling address
5. Detector supply voltage
6. Due time in format YYYYMMDD