© 2018 All rights reserved. Data subject to change without notice.
41
Rev. B | 2018.08
LPT-P- Operation Manual
NOTE:
The programming buttons inside the enclosure can be used instead of the magnetic wand,
if preferred.
7.2 Calibrating the Internal CO and/or NO
2
Sensor(s)
NOTE
: If an inappropriate concentration of span gas is applied during calibration, calibration may
succeed but it does not mean the equipment has been calibrated properly. CETCI is not responsible
for improperly calibrated transmitters. Follow manual instructions carefully.
The steps below show how to calibrate a CO sensor. Follow the same step for calibrating
an NO
2
sensor, choosing the menu options for the channel assigned to NO
2
.
To calibrate the internal sensor(s), the user must go through the following steps:
Step 1
In the Sensor Confi guration Menu (passcode 3032), press either ARROW key to scroll through the
menu and select the channel you want calibrate. Press ENTER and use the ARROW key to scroll to
Calibrate Zero. Press ENTER and use the ARROW keys to enter a value of 0.
Attach the regulator to the cylinder of zero air, insert calibration adapter into the sensor opening on
the front of the enclosure door (see Section
7.3.1 Calibration Adapter Plug
if the adapter will not fi t).
Press Enter to start the Zero calibration. The display will show a countdown from 18 seconds.