Press ENTER and use the ARROW keys to scroll through the menu and choose TEST. Press ENTER and
use the ARROW key to increase or decrease the value and the ENTER key to move to the next digit.
NOTE:
You will only be able to enter a value that falls within the factory confi gured range of the
sensor. If you try to enter a number outside that limit, the value will return to a valid value.
When fi nished, press ENTER to accept and Exit the screen. The gas level reading for that channel will
be immediately sent to the controller. This test reading will override any actual gas reading for that
channel for the duration of the test.
To stop the test, press ENTER to Exit. If you do not exit the test, the test will timeout after 5 minutes