B-19
When Line 4 indicates YES, press the TRANSMIT button to turn the beam on. Provided no fault
occurs, Lines 1 and 2 will appear as follows (Lines 3 and 4 will display the selected meters):
L O C A L 5 : 2 7 A
B E A M O N S E Q U E N C E P S T
Once the Beam On Sequence is complete, Line 2 will display the following:
H I G H V O L T A G E O N P S T
To exit this test mode, return to the HVPS TEST MODE menu. Use the arrow keys to choose
NO. The display (Lines 3 and 4) will return to the following:
H V P S T E S T M O D E
H V P S T E S T : N O
B.13.8 User
Tests
User Tests includes miscellaneous hardware tests available to the user. See Table B-7.
Table B-7. User Tests and Results
Function Name
Possible Entries/Results
Description
Lamp Test
Press ENTER to perform test. No
result is reported.
Illuminates all luminescent devices on the
panel for 15 seconds.
Beeper
Use arrow keys to enable or
disable beeper.
Press ENTER to start (same as other 2
loopback tests.)
Aux Serial Port back
test
Press ENTER to start; will display Loop
PASS if unit sees itself, FAIL if it doesn’t.
Press ENTER again or leave this item to
stop the test. For the unit to see itself, loop
TX+ to RX+ and TX- to RX- (if using RS-
422/485) or TX to RX (if using RS-232)
Line 3 will indicate:
U S E R T E S T
T
Line 4 will indicate (in order):
L A M P T E S T : E N T E R
If a test is performed, the result (as described above) will be reported in place of ENTER.
B E E P E R : E N A B L E
A U X L O O P B A C K : E N T E R