Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
26
Figure 25. Hold Error screen.
Hold Error
(Adjustable Audible Volume)
The
HOLD ERROR
screen, Figure 25, will appear if the pump has been inactive, without input, for
more than 10 minutes. See subsection “Hold” for description of
HOLDING
mode.
Press
CONTINUE
to return to the previous screen. Where settings can be adjusted, the pump
can be set to run immediately, or the pump can be set to run in a specified number of minutes. See
Figures 18A-18B to see the options on the
HOLDING
menu screen.
If the error cannot be resolved, press
POWER
to stop operation of the pump and put a different
pump into service.
Rotor Error
(Adjustable Audible Volume)
The
ROTOR ERROR
screen appears during
RUNNING
or
PRIMING
, when the pump has
Figure 26. Rotor Error screen.
Re-load the set tubing and
press CONTINUE.
detected an unusual operating condition for the rotor. See Figure 26. The
ROTOR ERROR
generally results from a problem with the pump set tubing around the rotor.
Check that the pump set is not damaged, and re-load the tubing around the rotor as described
in subsection “Loading Pump Sets”. Press
CONTINUE
to restart.
If the error cannot be resolved, press
POWER
to stop operation of the pump and put a
different pump into service.
Figure 27. Feed Error screen.
Feed Error
(Adjustable Audible Volume)
The
FEED ERROR
screen appears when the enteral formula is no longer being delivered because
the bag is empty or because of a clog between the pump and the bag. Check the feed line to find
the occlusion causing the blockage. If the occlusion cannot be cleared, load a new pump set, prime
it, and press
CONTINUE
to restart the feeding, Figure 27.
If the error still cannot be resolved, press
POWER
to stop operation of the pump and put a
different pump into service.
Flush Error
(Adjustable Audible Volume)
Figure 28. Flush Error screen.
The
FLUSH ERROR
screen appears when the flushing solution is no longer being delivered because
the flush bag is empty or because of a clog between the pump and flush bag. Check the flush line
to find the occlusion causing the blockage. If the occlusion cannot be cleared, load a new pump
set, prime it, and press
CONTINUE
to restart the feeding, Figure 28.
If the error still cannot be resolved, press
POWER
to stop operation of the pump and put a
different pump into service.
Figure 29. Flow Error screen.
Flow Error
(Adjustable Audible Volume)
The
FLOW ERROR
screen appears when the feeding solution is no longer being delivered because
of a clog between the pump and the patient. Check the line to find the occlusion causing the
blockage. If the error cannot be resolved, check the valve pocket in the pump set loading area for
dampness or dirt, which may be a cause for a false error. Clean and dry the valve pocket. If the
error still cannot be resolved, load a new pump set, prime it, and press
CONTINUE
to restart the
feeding, Figure 29.
If the error still cannot be resolved, press
POWER
to stop operation of the pump and put a
different pump into service.