![background image](http://html.mh-extra.com/html/covidien/kangaroo/kangaroo_operating-manual_2673891022.webp)
Table of Contents
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
18
Hold
Pump operation can be paused by pressing
HOLD
while it is
RUNNING
, such as in Figure 15A. The
status line will indicate
HOLDING
and the yellow LED indicator will be lit up to give a quick visual indication of an informational status in a darkened room.
Note:
the pump’s alarm will sound if it has been left
HOLDING
for 10 minutes without further input.
There are five options available during
HOLDING
. See Figures 18A and 18B.
Figure 18B. HOLDING mode
options, intermittent
feeding mode.
Figure 18A. HOLDING mode
options, continuous
feeding mode.
Clear Volume Fed
Press
CLEAR VOLUME
to re-zero the output feed totals “
mL FED
,” “
BOLUSES FED
,” and “
mL FLUSH
.”
Adjust Settings
Press
ADJUST SETTINGS
to return to one of the opening menus shown in Figures 7A-7D, thus allowing any setting to be
modified. If the pump set has been removed after going into
HOLD
, the opening menu will show
LOAD A SET
, as seen in
Figures 5A-5B.
Resume In __ Minutes
(Not Available in EZ Pump Mode)
Press “
RESUME IN __ MIN
” once to immediately place the Kangaroo Joey enteral feed and flush pump with pole clamp into a
30-minute resume running mode. Pressing the
RESUME IN __ MIN
button again will cyle the “resume in” time back to zero. No
other input is needed to start the countdown, as the time immediately counts down after the 30-minute value is reached.
Run
Press
RUN
to immediately restart pump operation. This option will not be available if the pump set is removed during
HOLD
.
History
(Not Available in EZ Pump Mode)
Press
HISTORY
to access the
HISTORY
screen. Use this screen to review feeding and flushing history. See subsection “More
Options” below for further information.