Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
15
Intermittent Mode — Adjust Bolus
(Not Available in EZ Pump Mode)
Figure 13A. ADJUST BOLUS menu.
Figure 13B. Setting the Bolus Rate.
Figure 13C. Setting the Number
of Boluses.
Figure 13D. Setting the volume
per Bolus.
Figure 13E. Setting the time
interval between
starts of bolus.
Intermittent Mode Feeding
For
INTERMITTENT
feeding mode, the
ADJUST BOLUS
option is available in the opening
menus (Figure 7C for a feed-only set or Figure 7D for a feed & flush set). Select this option to
program the feeding parameters from the
ADJUST BOLUS
screen, Figure 13A.
Press
BOLUS RATE
in Figure 13A to define the rate of delivery.
Press the buttons in Figure 13B to program the delivery rate from 1 to 400 in increments of
1 mL/hr. Press
ENTER
(top button) to exit the menu.
Press
NUMBER OF BOLUSES
in Figure 13A to define the total number of boluses to
be delivered.
Press the buttons in Figure 13C to program the number of boluses, up to a maximum of 99.
Press
ENTER
(top button) to exit the menu. The number of boluses can be set as zero, in
which case the pump will run until the supply is exhausted.
Press
VOLUME PER BOLUS
in Figure 13A to define the volume to be delivered in
each bolus.
Press the buttons in Figure 13D to program the volume per bolus from 1 to 3000 mL in
increments of 1 mL.
Press
ENTER
(top button) to exit the menu.
Press
BOLUS INTERVAL
in Figure 13A to define the time interval between starts of
bolus deliveries.
Press the buttons in Figure 13E to program the time interval between starts of bolus
deliveries from 1 to 24 hours in increments of 1 hour. Press
ENTER
(top button) to exit
the menu.
The pump will automatically limit bolus settings which exceed pump delivery capabilities.
For example, a bolus delivery volume of 200 mL cannot be made once per hour if a rate of 150
mL/hr is already programmed. In this case, the maximum volume allowed would be 150 mL.
Note:
If the
NUMBER OF BOLUSES
is reprogrammed during a feeding (from HOLD mode),
the pump will restart the bolus feeding and will deliver all the boluses as defined by the
new setting.
Intermittent Mode Feeding – Bolus Max
(Not Available in EZ Pump Mode)
The
BOLUS MAX
option can be used to deliver a bolus at a high rate, similar to
gravity feeding.
Press
BOLUS MAX
in the
BOLUS RATE
menu, Figure 13B, to program the
BOLUS
MAX
mode.
The feeding rate for
BOLUS MAX
is 999 mL/hr.