Table of Contents
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
19
Flush Now
(Not Available in EZ Pump Mode)
During the
RUNNING
mode, the pump can be diverted to immediately perform a flush (flush on demand) when a feed/flush set is
loaded by pressing
FLUSH NOW
in the
RUNNING
menu, input a desired flush volume, then press
ENTER
. See Figure 15A.
When
FLUSH NOW
is pressed, the screen will prompt for the volume to be immediately flushed, and will show the previously
programmed flush volume as default but may be adjusted to the users desired volume. The screen will appear as the standard Flush
input screen as shown in Figure 12B (or 14B). Setting the
FLUSH NOW
flushing volume is temporary and will not change the main/
periodic flushing volume previously programmed. If
FLUSH NOW
is pressed by accident; the user may either press the cancel
button or allow the screen to return to the
RUNNING
screen after 10 seconds, if there is no input for the flush volume.
If stopping a
FLUSH NOW
before the set volume has been delivered is desired, press
FLUSH NOW
and enter a volume of zero and
press
ENTER
. The
FLUSH NOW
will stop immediately and the pump will return to the
RUNNING
mode.
More Options
Figure 19. MORE OPTIONS menu.
Press
MORE
in one of the opening menus, Figures 5A-5B or Figures 7A-7D, to access the
MORE OPTIONS
screen, Figure 19.
Buzzer
Press
BUZZER
in the
MORE OPTIONS
menu to access the
BUZZER
screen and increase or
decrease the alarm buzzer volume. The alarm will be audible as the volume setting is changed.
This sets the volume level for any warning or error that may occur with the pump. Critical errors of
System Error and Pump Set Dislodged will always use the maximum volume setting.
History
(Not Available in EZ Pump Mode)
Figure 20. Feed and Flush totals
for up to 72 hours of previous
history, excluding data from the
current hour.
Press
HISTORY
in the
MORE OPTIONS
menu to access the
HISTORY
screen. Press the
buttons in Figure 20 to set the time history of interest. The totals of
FEED mL
and
FLUSH mL
will be shown for the requested history time.
Warning:
A maximum of 72 hours of past history can be retrieved, excluding data from the
current hour. Note: The 72-hour history includes time when the pump is powered Off.
Languages
Press
LANGUAGES
in the
MORE OPTIONS
menu to change the preferred language for the
screen interface.
Continuous/Intermittent
(Not Available in EZ Pump Mode)
Press
CONTINUOUS/INTERMITTENT
in the
MORE OPTIONS
menu to access the
SELECT MODE
menu, Figure 10B. Use this
menu to select between continuous feeding mode and intermittent feeding mode. See subsection “Selecting Continuous or
Intermittent Feed Mode” for more information.
Press
DONE
in the
MORE OPTIONS
menu to exit the menu.