Covaris AFA ultraCUBE User Manual Download Page 7

 

 
 
AFA ultraCUBE User Manual 

 

1.3 

For the safety of operating personnel 

 

1.  Make sure that the power supply module is properly grounded.  Employ a power cable 

with a grounding prong, plugged to a power receptacle with a grounded receptacle.  DO 
NOT operate the unit if it is not properly grounded.  

2.  The unit is equipped with a power plug appropriate for the destination country.  DO NOT, 

under any circumstances, remove the grounding prong from the power cord. 

3.  Always disconnect power to the equipment (unplug) before performing any service or 

maintenance.  Reconnect power only as instructed. 

4.  Employ only the 12 volt DC power module provided with the unit, or a replacement 

provided by Covaris. 

5.  Follow all precautions when changing the UV lamp.  Avoid exposure to direct or strongly 

reflected germicidal ultraviolet rays. Germicidal ultraviolet rays are harmful to the eyes 
and skin. 

6.  THE ULTRAVIOLET LAMP CONTAINS MERCURY (Hg).  Handle carefully to avoid 

breaking the lamp.  Dispose of used lamps according to disposal laws.  For further 
information, see one of the following websites or contact Covaris. 

www.lamprecycle.org

 (North America) 

www.lamprecycling.com

 (North America) 

www.mercuryrecycling.co.uk

 (UK) 

www.malampe.org

 (France) 

7.  If there is a safety related malfunction, DO NOT operate the equipment and contact 

Covaris immediately.  

8.  If the equipment is used in a manner not specified by the manufacturer, the protection 

provided by the equipment may be impaired. Do not remove safety labeling. 

  

1.4 

To prevent damage to the equipment:  

 

1.  The AFA ultraCUBE system is designed for indoor use and to operate in typical ambient 

laboratory conditions e.g., 19°C to 25°C  (66°F to 77°C)

.  

DO NOT operate the instrument 

in a cold room environment.  The system is designed to operate with a water bath and re-
circulating heater/chiller apparatus to control sample temperature.  

2.  The AFA ultraCUBE system is designed for use with chilled or room temperature bath 

water.  Condensation, however, may form on interior surfaces and drip onto the bench 
top when the instrument bath water is chilled. 

3.  Distilled or de-ionized water should be used to fill the water bath.    
4.  Do not employ isopropyl alcohol, ammonia-based or abrasive cleaners on the water tank, 

as these will damage the acrylic surfaces. 

5.  When bleaching the Ultrasonicator system, turn off and disconnect the hoses to the AFA 

ultraCUBE system.  

6.  The WCS 2.0 requires periodic filter and UV lamp replacement, as described in this 

manual. 

7.  Do not directly handle the UV lamp 

– only handle it while wearing cotton or laboratory 

gloves to avoid fingerprints on the lamp glass. 

 

1.5 

Acronyms employed in this User Manual 

 

WCS 2.0 

– Water Conditioning System version 2 

UV 

– Ultraviolet 

AFA™ – Adaptive Focused Acoustics 

 

 

Summary of Contents for AFA ultraCUBE

Page 1: ...traCUBE User Manual 1 AFA ultraCUBE Water Management System USER MANUAL A system developed for maintaining bath water quality in Covaris AFA instrumentation Product Name AFA ultraCUBE Product Number 5...

Page 2: ...NIVERSAL PRECAUTIONS Universal Precautions should be followed on all specimen samples regardless of whether a sample is known to contain an infectious agent Laboratories handling specimen samples are...

Page 3: ...D YOU HAVE ANY QUESTIONS CONCERNING THIS EQUIPMENT Warranty Services The purchased equipment is covered by a twelve 12 month warranty which includes all the service and support necessary so that the c...

Page 4: ...AFA ultraCUBE User Manual 4 This page intentionally left blank...

Page 5: ...ENTS 1 0 INTRODUCTION 6 2 0 ultraCUBE Installation and Setup 10 3 0 Indicators and Controls 12 4 0 System Operation 14 5 0 Specifications 15 6 0 Service 16 Appendix B User Guidelines and Troubleshooti...

Page 6: ...enience if disregarded CAUTION Equipment damage may occur WARNING Personal injury may occur 1 2 Safety Symbols and Warnings The following symbols are employed in product labeling and in this manual Wh...

Page 7: ...T operate the equipment and contact Covaris immediately 8 If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Do not remove...

Page 8: ...cally switch between chilling heating mode and standby mode in which water continues to circulate through the WCS to maintain water quality without the energy consumption required for cooling System C...

Page 9: ...continuously maintaining water quality and relieving the need for daily water changes in Covaris Adaptive Focused Acoustics AFA systems Figure 1 The Covaris Water Conditioning System The WCS 2 0 incor...

Page 10: ...ermoCube manual on CD Setup Refer to Appendix A of this manual to install the quartz tube and UV lamp into the WCS 2 0 Place the WCS 2 0 and the CH05 on the bench next to the Ultrasonicator They may b...

Page 11: ...AFA ultraCUBE User Manual 11...

Page 12: ...or 60 Hz NOTE The WCS 2 0 unit is shipped with the UV chamber quartz tube and UV lamp not installed Install these items before powering the unit or using it in any way See Appendix A for UV lamp and...

Page 13: ...ractical after the MAINTENANCE indicator illuminates The indicator is shut off by pressing an internal RESET button once replacement is performed Please see Appendix A for UV lamp and filter replaceme...

Page 14: ...is is normal and the turbulence subsides once the WCS 2 0 fills with water The CH05 and WCS 2 0 system when operating will automatically circulate bath water filter it and retard the growth of algae o...

Page 15: ...gh 33 cm 13 inches deep 28 cm 11 inches wide and 33 cm 13 inches high Weight Approximately 4 5 kilograms 10 lbs Approximately 12 7 kilograms 28 lbs Ambient Temp 19 C to 25 C 66 F to 77 C 0 C to 40 C 3...

Page 16: ...ice covarisinc com CAUTION PREPARING THE WCS 2 0 FOR SHIPMENT PURGE THE WCS 2 0 OF WATER AND REMOVE THE UV LAMP QUARTZ TUBE Please purge the WCS 2 0 of standing water prior to return shipment particul...

Page 17: ...2 0 500575 LE Series CH05 Hose Repair Kit 500576 LE Series WCS 2 0 Hose Repair Kit Quartz Tube Installation or Replacement The Ultraviolet UV chamber consists of the outer chamber in which a quartz tu...

Page 18: ...and draw it out of the UV chamber by gently pulling upward There are two O rings which will likely come out with the tube If the quartz tube is being installed for the first time the two O rings will...

Page 19: ...the water filter UV Lamp Enclosure Cap and Cable 5 Gently pull on the cable to withdraw the lamp socket from the enclosure If a lamp is already installed in the enclosure it will be drawn out with th...

Page 20: ...ocessing of infectious or hazardous biological samples laboratory gloves should be worn and the used filter cartridge treated as biohazard waste in case any biological sample accidentally entered the...

Page 21: ...ideways to release the fitting Repeat for the upper fitting The quick release fittings will seal and prevent water from draining from the cartridge or tubing 3 Pull on the used filter cartridge to rem...

Page 22: ...ensure there are no water leaks 7 Once the system is verified to be operating correctly press and hold the RESET button located on the side of the electronics enclosure for about 5 seconds This action...

Page 23: ...fittings No operation no power No 12V supply to the WCS 2 0 Verify that power pack is plugged into AC wall power and connected to the 12V jack on the rear panel of the WCS 2 0 MAINTENANCE indicator is...

Reviews: