AFA ultraCUBE User Manual
21
1. Note the orientation of the tubes and filter cartridge on
the back wall of the open WCS 2.0 cabinet, next to the
UV chamber. The flow arrows should point towards
the bottom of the WCS 2.0 cabinet.
2.
Position a paper towel under the lower
Quick Connect fitting on the filter cartridge.
Push the locking tab sideways to release the
fitting. Repeat for the upper fitting. The quick
release fittings will seal and prevent water
from draining from the cartridge or tubing.
3. Pull on the used filter cartridge to remove it from the plastic grips.
4.
Install the replacement filter by pressing it down
into the plastic grips, taking care that the flow arrows
point towards the bottom of the WCS 2.0 cabinet.
Reconnect the fittings as originally found by pressing
them into each Quick Connect socket until they snap
securely.
Push tab to
release the fitting.