AFA ultraCUBE User Manual
22
WARNING:
Moderate pinch hazard - Do not place your fingers under the filter
cartridge near the retaining clips while pressing the cartridge into place.
Inspect and Prepare the WCS 2.0 for Use
5. With the WCS 2.0 cover off, verify that the UV lamp enclosure is fully closed (the
black grommet covers the opening
and the UV lamp is not visible). Fill the Covaris
bath with distilled or de-ionized water, and reconnect the 12V power module. Turn
ON the WCS 2.0 and CH05.
6. Verify that bubbles briefly appear in the bath as air is purged from the WCS 2.0 filter.
Inspect fittings to ensure there are no water leaks.
7. Once the system is verified to be operating correctly,
press and hold the RESET button located on the side of
the electronics enclosure for about 5 seconds. This
action turns the MAINTENANCE indicator off. (It will turn
on again at the next multiple of 9500 hours.) Verify that
the MAINTENANCE indicator turns off and remains off.
NOTE: The MAINTENANCE indicator will remain illuminated if the UV lamp
has not been correctly plugged in or is not operating.
8. Turn power OFF and disconnect the 12V power module.
9. Replace the WCS 2.0 cover. Reinstall 6 screws at the rear of the cover, and reinstall
the 2 screws at the bottom front of the cover. Hand-tighten to secure each screw.
10. Reconnect the 12V power module. Restart the WCS 2.0. Add water as necessary to
the water bath.