• Do not allow patients to interact with the C3
+
, unless when directly instructed by a healthcare
professional.
• Do not touch the electrode connections while the USB cover is removed.
• Do not give the C3
+
to a user or patient, without the USB cover being properly closed.
• Do not touch patient and C3
+
simultaneously while C3
+
is charging.
• Do not use the C3
+
during MRI scans.
• Do not use the C3
+
with a defibrillator.
• The C3
+
cannot detect pacemaker pulses.
• Do not expose the device to strong sources of static electricity or electromagnetic fields.
• Do not leave C3
+
on top of or next to other electrical equipment.
• Do not use C3
+
with cables different than the one provided by Cortrium.
• Do not submerge the C3
+
in liquid.
• Do not clean the C3
+
with agents other than those listed in the cleaning instructions in this
manual.
• Do not damage the C3
+
through drops, violent shaking or crushing.
• Do not use the C3
+
on patients with highly sensitive skin or known skin-related allergies.
• Do not use the C3
+
on breached skin.
• Do not use the C3
+
on patients below 10 kilos of body weight.
• The C3
+
is not a toy. Usage on children should be under strict supervision of adults.
• Do not put C3
+
into mouth under any circumstances.
• Do not alter the C3
+
. Any modification of the C3
+
is strictly prohibited.