background image

6

7

CORNING LSE MULTI PURPOSE DIGITAL VORTEXER

Instruction Manual

3.4 Environmental Requirements

The Corning® LSE™ Multipurpose Digital Vortexer has been built for operating in both 
cold and heated laboratory environments and in high humidity environments as follows:

At temperatures between 0°C and +65°C

In humidity up to 85% (without condensation)

4.0 Installation of Vortexer Platforms 

Cat. No. 

Platform Descriptions 

48010 

Platform for 4 microplates 

48011 

Platform with non-slip rubber mat 

48012 

Spring loaded platform for flasks, bottles, or tube racks

The Corning LSE Multipurpose Digital Vortexer requires the installation of the selected 
platform. Please follow the instructions below for installing the platform.

Before you attach the platform, make sure that the slide stabilizers are properly 
fastened, adjusted, and lubricated with the included grease.

The flat mat platform has a circle cut-out in the rubber and screw holes in the center. 
Remove the cut-out rubber circle and screw down the 4 screws onto the Vortexer. After 
attaching the platform, remove the non-adhesive backing paper on the rubber circle, 
and paste it back into place.

The grease is used to lubricate the slide stabilizers. It is sufficient for 1,000 working 
hours (approx. 1 year). After that time you should remove the platform and clean, 
adjust, and re-grease the slide stabilizers. If they are worn out, you should replace them 
with new stabilizers and set to appropriate height.

5.0 Operating Instructions

5.1 Introduction

1. POWER switch– switches Vortexer ON (light on) or OFF.
2. RUN green signal light – lit when the Vortexer is running.
3. TIME yellow signal light – lit when the Vortexer is set on time.
4. RPM yellow signal light – lit when the Vortexer is set on rpm.
5. ENCODER – by rotating the encoder right (+) or left (-) you can change the set TIME or 

RPM. Push the encoder to switch between TIME and RPM set values. If you rotate the 
ENCODER knob fast, the values go up or down on the display very quickly.

6. START/STOP button – press to START or STOP operation.

5.2 Basic Operation

Press the POWER switch on the control panel. The Vortexer automatically detects 
supply frequency of 50 or 60Hz, and displays F50 or F60. Then the display switches to 
show the set TIME and illuminates the TIME light.

Power

Run

Speed (RPMx10) time (min)

Time

Speed

Start

Stop

PUSH

(Time/Speed)

ENCODER

1

2

3

4

5

6

Power

Run

Speed (RPMx10) time (min)

Time

Speed

Start

Stop

PUSH

(Time/Speed)

ENCODER

Grease

Sliding Stabilizers

0.7 mm

Summary of Contents for LSE 6709

Page 1: ...Corning LSE Multipurpose Digital Vortexer Instruction Manual Cat No 6709 115V Cat No 6710 230V ...

Page 2: ...ations 4 3 0 Installation 5 3 1 Unpacking 5 3 2 Selecting a Suitable Location 5 3 3 Attaching the Power Cord 5 3 4 Environmental Requirements 6 4 0 Installation of Vortexer Platforms 6 5 0 Operating Instructions 7 5 1 Introduction 7 5 2 Basic Operation 7 5 3 Additional Operations 11 6 0 General Precautions 12 7 0 Troubleshooting Guide 12 8 0 Maintenance and Cleaning 12 9 0 Limited Warranty 13 10 0...

Page 3: ...sing parts Open the box and lift the device together with inner packing out of the box Remove the inner packing and check that the Vortexer has not been visibly damaged during transportation Keep the packing material until you are sure that the Vortexer works properly Check information on the rear data label and verify the following are correct Model Serial number Electrical rating Should any kind...

Page 4: ...ack into place The grease is used to lubricate the slide stabilizers It is sufficient for 1 000 working hours approx 1 year After that time you should remove the platform and clean adjust and re grease the slide stabilizers If they are worn out you should replace them with new stabilizers and set to appropriate height 5 0 Operating Instructions 5 1 Introduction 1 POWER switch switches Vortexer ON ...

Page 5: ...e RPM signal light is on By rotating the encoder right or left the desired RPM can be selected 34 340 rpm 120 1 200 rpm Press the START STOP button to start operation Power Run Speed RPMx10 time min Time Speed Start Stop PUSH Time Speed ENCODER Power Run Speed RPMx10 time min Time Speed Start Stop PUSH Time Speed ENCODER Power Run Speed RPMx10 time min Time Speed Start Stop PUSH Time Speed ENCODER...

Page 6: ... RPM signal light must be on If you want to view the set value for TIME during vortexing rotate the encoder ONE CLICK right or left The TIME signal light will flash for 2 seconds and the display will show the set TIME After 2 seconds the display will return to the count TIME and the signal light will stop flashing NOTE The Vortexer must be in the TIME mode the TIME signal light must be on Power Ru...

Page 7: ...OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be to repair or replace at its option any product or part thereof that proves defective in material or workmanship within the warranty period provided the purchaser notifies Corning of any such defect Corning is not liable for any incidental or consequential damages commercial loss or any other damages from the use of this prod...

Page 8: ...ste electrical and electronic equipment WEEE the Corning LSE Multipurpose Digital Vortexer is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment WEEE provided with the products and available at the following link www corning com weee ...

Page 9: ... or contact your local Corning sales office 2018 Corning Incorporated All rights reserved 5 18 CLS AN 498 11319 30544 Corning Incorporated Life Sciences 836 North St Building 300 Suite 3401 Tewksbury MA 01876 t 800 492 1110 t 978 442 2200 f 978 442 2476 www corning com lifesciences ASIA PACIFIC Australia New Zealand t 61 427286832 China t 86 21 3338 4338 f 86 21 3338 4300 India t 91 124 4604000 f ...