C O R N I N G   C E L L   C O U N T E R

Instruction Manual

Live/Dead Sample Loading

To perform a live/dead analysis, we recommend using the standard Trypan blue 
technique. Mix a minimum of 20 µL sample with 20 µL trypan blue by using a 20 µL 
pipet. Next load 10 µL of the sample mixture into the counting chamber.

Note: Remember to correct the results when a dilution is used, as the Corning®  
Cell Counter does not correct for this automatically.

Counting Chamber Placement

The optical window of the Corning Cell Counter must be clean. You can check this by 
monitoring the live screen. If any debris is present, you will see it in the CytoSMART™ 
Cloud App.

If any debris is visible, clean the optical window using lint-free wipes before placing the 
counting chamber in the adapter plate.

When the sample is loaded in the counting chamber and the adapter plate has been 
installed, you can place the counting chamber in the adapter plate that has been 
inserted on the Corning Cell Counter. 

4.0 Performing a Cell Count

After placing the loaded counting chamber on the Corning® Cell Counter, you can see an 
image of the cell sample in the live screen (Figure 17).

Focusing of the Cells

Use the focus knob to adjust the focus of the image. To focus, turn the knob clockwise or 
counter clockwise. For best results, ensure you have the correct focus on your tablet or PC.

For proper analysis, you should be seeing a contrast: living cells having bright centers 
and dark edges (Figure 18). To achieve this, using the focus knob, set the focus just 
above the cell.

We recommend trying multiple focus levels to determine which focus setting to use.

Figure 18.

 Correct focus settings

Figure 17. 

Live screen of the CytoSMART™ 

Cloud App

A

Important: 

The image seen on the live screen is a digital magnification of the field of 

view, to allow for better focusing. The unmagnified image can be seen by clicking on the 
magnifying glass (Figure 17A).

Summary of Contents for 6749

Page 1: ...Corning Cell Counter Accurate Cell Counts in a Flash Instruction Manual Cat No 6749...

Page 2: ...Plate Placement 12 Sample Loading 13 Live Dead Sample Loading 14 Counting Chamber Placement 14 4 0 Performing a Cell Count 15 Focusing of the Cells 15 Performing a Count 16 5 0 Connecting to the Clou...

Page 3: ...the device 1 0 Introduction Technology to Make Your Work Easier Thank you for purchasing the Corning Cell Counter The Corning Cell Counter is a benchtop automated image based cell counter which uses...

Page 4: ...ew of the Corning Cell Counter 1 Light Arc 2 USB Cable 5 Focus Knob 3 Optical Window 4 Adapter Plate 2 0 Getting Started Product contents include the following items Figure 2 1 Manual and Quick Starte...

Page 5: ...on has completed an icon for the CytoSMART Cloud App will be visible on the desktop Figure 3 4 Start the CytoSMART Cloud App by double clicking the icon on the desktop A notification to connect the de...

Page 6: ...can perform a single count or multiple counts within a project The CytoSMART Cloud App software automatically stores the results within a project which you can view later in the CytoSMART Connect Figu...

Page 7: ...ownwards until it clicks into place on top of the Corning Cell Counter Figure 15 Figure 13 Start cell count Figure 14 Slide the adapter plate into place Figure 15 Carefully press the adapter plate dow...

Page 8: ...installed you can place the counting chamber in the adapter plate that has been inserted on the Corning Cell Counter 4 0 Performing a Cell Count After placing the loaded counting chamber on the Corni...

Page 9: ...nnect environment 5 If you want to perform another cell count simply press Start new cell count E Figure 19 Overview of the result page A B C D E 5 0 Connecting to the Cloud One of the features of the...

Page 10: ...verview of all projects Figure 23 Navigating the portal is simple If you have multiple projects in an account use the Search Projects bar A to quickly search for a project To create a new project sele...

Page 11: ...circles Experiment Pane The experiment pane provides an overview of the experiment i e project start date owner number of cells and viability Notes Pane Notes that you have added after counting the ce...

Page 12: ...r before loading a new sample Figure 29 Cell clump example Figure 30 Air bubble example Figure 31 Debris present in the counting chamber No cells are visible This may be due to cell concentration of t...

Page 13: ...causes not arising from defects in original material or workmanship This warranty does not cover motor brushes fuses light bulbs batteries or damage to paint or finish Claims for transit damage shoul...

Page 14: ...te electrical and electronic equipment WEEE the Corning Cell Counter is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the...

Page 15: ...tact your local Corning sales office 2018 Corning Incorporated All rights reserved 2 18 CLS EQ 486 11319 30504 Corning Incorporated Life Sciences 836 North St Building 300 Suite 3401 Tewksbury MA 0187...

Reviews: