Turn on water supply and main power switch (located on top of electrical box). Make the following system
checks:
NOTE:
If unit will not start be sure water reservoir is full. Low water safety control must be properly adjusted
to start and shut down unit. If water level drops below bottom of reservoir, unit must shut down. Adjustment is
made by moving magnet up or down.
FIGURE 1. WATER RESERVOIR
Water Level – If necessary adjust Float by bending float arm up or down as needed, push float assembly down
until unit stops running. Release float and unit will restart. Keep water in reservoir at level line while unit is in
operation. See Figure 1.
Low Water Safety Control – Adjust magnet by bending magnet arm as shown in figure 1 to shut down unit if
the water level drops below the line on the side of the reservoir.
Bin Control – Remove four screws from top of bin cover and lift cover so bin control plate can be manually
lifted until unit shuts down. Release plate and unit will restart (NOTE: the dispense button must be depressed
for 45 seconds before unit will start). Replace screws.
Dispense Switch and Mechanism – By depressing the dispense switch, the dispense mechanism door on the
storage bin will open, and agitator will rotate counterclockwise.
NOTE:
If any of these checks or adjustments cannot be achieved, refer to Troubleshooting Section of this
manual or call our technical support center for assistance.