3) Output On Time Adjustment
Press FUN and NEXT (2
times
). PO.t will display.
3.5 sec is factory default set value. Press UP/Down or type numeric value up
to 99.9 seconds maximum. Press FUN to store this set value.
Press FUN and NEXT (3
times
). Ne.t will display.
3.5 sec is factory default set value. Press UP/Down or type numeric value up
to 99.9 seconds maximum. Press FUN to store this set value.
4) Output Voltage Adjustment
Press FUN and NEXT (4
times
). Po.V will display.
4.5 kV is factory default set value. Press UP/Down or type numeric value up
to 7.0 kV maximum. Press FUN to store this set value.
Press FUN and NEXT (5
times
). Ne.V will display.
3.9 kV is factory default set value. Press UP/DOWN or type numeric value up
to 7.0 kV maximum. Press FUN to store this set value.
7 Setup and Operation
5) Cleaning Cycle Alarm
Press FUN and NEXT (6
times
). CA.O will display.
Disabled is factory default set value. Press UP/Down to change enable or
disable cleaning cycle alarm activation. Press FUN to store this set value.