D5.4: Multi-sensor system integrated into wearable fabrics
36/42
MOBISERV
FP7 248434
Table 19: – Electromagnetic immunity –
IMMUNITY test
IEC 60601
test level
Compliance
level
Electromagnetic environment
– guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
N/A
Mains power quality should be
that of a typical commercial or
hospital environment
Voltage dips,
short interruption
and voltage
variations on
power supply
lines
IEC61000-4-11
N/A
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the MOBISERV data logger
requires continued operation
during power mains
interruptions, it is recommended
that the MOBISERV data logger
be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Portable and mobile RF communications equipment should be used no closer to any part of
the MOBISERV data logger, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation
distance
Conducted
RF
IEC 61000-4-5
N/A