D5.4: Multi-sensor system integrated into wearable fabrics
19/42
MOBISERV
FP7 248434
ANT
TM
Channel Configuration
(pairing bit disabled)
Transmission Type
1
No shared address, Global
Data Identification Byte
present
Network configurations
Channel type
0x10
Bidirectional Master Transmit
Channel
Channel Period
16384
2 Hz Message Rate
Network number
0
Default public network
Message Rate
2 Hz
Message Rate
RF Frequency Offset
0 MHz for module 1
72 MHz for module 2
Default Frequency 2400 MHz
Default Frequency 2472 MHz
3.2.4
Data Format
3 different data channels, more precisely the cuff sensor, abdominal sensor and battery
voltage, are assigned to the physical ANT channel specified by Device ID parameters listed
in Table 3. The data are transmitted according to ANT SensRcore™ Global Data Message
Format that is reported in Table 4, in which Global Data identification byte value is 0x41
(analogue data).
Table 4 Global Data Message Format
Data 0
Data 1
Data 2
Data 3
Data 4
Data 5
Data 6
Data 7
Global Data
Identification
Byte
Global
Defined
Data 1
Global
Defined
Data 2
Global
Defined
Data 3
Global
Defined
Data 4
Global
Defined
Data 5
Global
Defined
Data 6
Global
Defined
Data 7
Table 5 shows the significance of transmitted bytes:
Table 5 ANT
TM
Message Payload
ANT Message
Payload Byte
Parameters
Type
Range
Description
Global Data Byte 3
Analogue Signal Source
UCHAR
0..255
Indicates signal source:
0x00 --> Temp sensor 1
(hip sensor)
0x01: Temp sensor 2
(cuff sensor)
0xFF: Battery Voltage
Global Data Byte 4
Max (8-bit)
UCHAR
0..255
8 most significant bits of
the maximum recorded
analogue level