background image

HTML Access (IR data structure)

The structure for the IR information: 
typedef struct lepton_settings_struct 

  __packed uint16_t isotherm_min_enable; 
  __packed uint16_t isotherm_max_enable; 
  __packed uint16_t isotherm_min; 
  __packed uint16_t isotherm_max; 
  __packed uint16_t palette; 
  __packed uint16_t units; 
  __packed uint16_t led; 
  __packed uint16_t reserved1; 
  __packed char name[16]; 
  __packed char serial[STRING_LENGTH]; 
  __packed enum View_Mode view_mode; 
  __packed enum DateFormat date_format; 
  __packed enum TimeFormat time_format; 
  __packed enum Dhcp dhcp; 
  __packed uint32_t static_ip_addr; 
  __packed uint32_t static_netmask; 
  __packed uint32_t static_gateway; 
  __packed uint32_t crc; 
} lepton_settings_type;

typedef struct 

lepton_info_struct 


  __packed uint16_t bin_max; 
  __packed uint16_t bin_min; 
  __packed uint16_t max_x; 
  __packed uint16_t max_y; 
  __packed uint16_t min_x; 
  __packed uint16_t min_y; 
  __packed uint16_t palette_depth; 
  __packed int16_t ambient; 
  __packed uint16_t palette_lut[PALETTE_
    LARGEST]; 
  __packed uint16_t palette_temp[PALETTE_
    LARGEST]; 
  __packed lepton_settings_type settings; 
  __packed uint32_t firmware; 
  __packed uint32_t alarm_status; 
  __packed uint16_t token; 
} lepton_info_type;

typedef struct

 alarm_settings_struct 


__packed uint16_t alarm[25]; 
__packed uint32_t alarm_enable; 
__packed uint16_t token; 
} alarm_settings_type;

enum 

Palette { GREY_POSITIVE=0, GREY_

NEGATIVE, IRON, HOTMETAL, RAINBOW, AMBER, 
SEPIA }; 
 

enum

 View_Mode { ISOTHERM_OVERLAY, 

THERMAL_ONLY, VISIBLE_ONLY, HIGHLIGHT, 
MAX_VIEW_MODE }; 

enum

 Units { CELSIUS, FARENHEIT, KELVIN, 

MAX_UNITS }; 

enum

 DateFormat { YYYYMMDD, DDMMYYYY, 

MMDDYYYY, MAX_DATE_FORMAT }; 

enum

 TimeFormat { HH24MMSS, HH12MMSS, 

MAX_TIME_FORMAT}; 

enum

 Dhcp { DHCP, STATIC_IP, MAX_DHCP};

 

enum

 Led { LED_DISABLED, LED_ENABLED, 

MAX_LED};
 
DateFormat and TimeFormat are not used.

palette_lut[PALETTE_LARGEST] is an array of 
colors 

Colours are RGB565, 16 bit colour. 

palette_temp[PALETTE_LARGEST] is an array

of temperatures in cK 

The temperatures of each palette colour.

RUGGED 

AUTHORITY

MONITIR Series Thermal Imaging Cameras

 – User Manual

31

Document Reference MONITIRSERIESUM Rev. A

Summary of Contents for MN4000 series

Page 1: ...USER MANUAL MN4000 MN4100 SERIES THERMAL IMAGING CAMERAS RUGGED AUTHORITY www cord ex com...

Page 2: ...rmal Imager CONTENTS Introduction 8 Quick Start Guide 10 Mechanical Installation 12 Important Network Information Parameters 13 Camera web server 14 CorDEX MONITIR software 15 MODBUS HTML 15 IP Addres...

Page 3: ...r communicating with devices such as Programmable Logic Controllers PLCs The CorDEX MN4xxx may be installed in a standalone or multi camera mode with each camera incorporating a built in web server fo...

Page 4: ...settings You have now accessed the CorDEX MN4100 series camera From this screen you can setup and provide basic control commands to the camera 9 Click Submit The camera will now restart with desired...

Page 5: ...the rear of the panel is clear of ribs or other obstructions prior to drilling 5 Using the drill bit specified on the template xxx drill the three fixing holes located around the perimeter 6 Using an...

Page 6: ...mation thermal imager is intended for permanent installation into fixed locations for industrial automation control and safety monitoring purposes Installation is via an M6x4 threaded boss located on...

Page 7: ...cket and power terminals 2 Insert power and communications cables into camera via 2xM16 compression glands provided NOTE these glands will accept cables with outside diameters from 3mm 7mm 3 Connect 1...

Page 8: ...K INFORMATION AND PARAMETERS IMPORTANT NOTICE CorDEX MN4000 cameras are shipped as DHCP enabled as setup configuration is via the USB port located on the front of the camera CorDEX MN4100 cameras are...

Page 9: ...9Hz Home tab The Home tab is the default tab after logging into the camera Under the Home tab it is possible to change the image type image colourisation set rising and falling isotherms and to toggl...

Page 10: ...lling Isotherm click the OFF ON toggle switch to the ON position green and insert the alarm value in the box adjacent to the button Click Send to Camera Any temperature in the cameras Field of View FO...

Page 11: ...ier input by the installer which pinpoints the camera location once set this should not be changed unless the camera duty is altered The unit of measure can be selected from three potential options Ce...

Page 12: ...of identifying where the camera is located in a particular facility User configurable 3 Ip Address The IP Address of the camera User configurable 4 Mac Address The Mac Address of the camera Factory se...

Page 13: ...ommended initial setup mode for MN4100 cameras as they are shipped with a factory set static IP address To communicate with the camera ensure the PC running the MONITIR Application is connected to the...

Page 14: ...codes 3 6 and 16 with the addresses as defined in the following table MODBUS Function code 3 6 16 TCP Registers HTML Browser A web browser may be used to setup all aspects of the camera operation the...

Page 15: ...re cfk is at s 8 2 176 5 dhcp is at s 8 2 176 25 upper 8 bits loc is at s 8 2 176 8 next 16 characters ethIpAddress is at s 8 2 176 26 next 4 bytes ethSubnetMask is at s 8 2 176 28 next 4 bytes ethGat...

Page 16: ...mbient __packed uint16_t palette_lut PALETTE_ LARGEST __packed uint16_t palette_temp PALETTE_ LARGEST __packed lepton_settings_type settings __packed uint32_t firmware __packed uint32_t alarm_status _...

Page 17: ...on as default MN4000 ship with DHCP Enabled non static address MN4100 Ship with static IP address of 192 168 1 100 Ip Address Discovery Auto Find IP The units use the network DHCP server to assign IP...

Page 18: ...Client ReceiveTimeout 3000 listBox1 Items Clear send data MonitIR client Send RequestData RequestData Length new IPEndPoint IPAddress Broadcast 46000 then receive data while true try var ServerRespon...

Page 19: ...locally at the camera MN4100 Factory reset The MN4100 Factory Reset is achieved by pressing and holding the Factory Reset button located at the rear of the camera for one second PB 15 14 HWVersion 1 0...

Page 20: ...R YOUR PRODUCT Register your product now at http www cord ex com register your product and receive a 12 month extended warranty Terms Conditions apply www cord ex com RUGGED AUTHORITY NOTES RUGGED AUT...

Page 21: ...Copyright 2015 CorDEX Instruments Limited All other brand and product names are trademarks of CorDEX Instruments Limited Document Reference MONITIRSERIESUM Rev A...

Reviews: