Appendix 1 Product Specifications
A1.1 Monitor
Physical
Characteristics
Dimensions:
347 mm x 330 mm x 126 mm (13.7
“
x 13
“
x 5
“
)
Weight:
about 6 kg (13 lbs)
Safety
Complies with:
IEC 60601-1:1988+A1+A2, EN 60601-1:1990+A1+A2, IEC/EN 61157,
IEC/EN 60601-2-37, IEC/EN 60601-1-2:2001+A1
Anti-electric Shock Type:
Class I equipment with internal power supply
Anti-electric Shock Degree:
FHR1, FHR2, TOCO, FM, FS B
IUP BF
DECG CF
Degree of Protection against Harmful Ingress of Water:
Ordinary equipment (sealed equipment without liquid proof)
Degree of Safety in Presence of Flammable Gases:
Equipment not suitable for use in presence of flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
Disinfection/Sterilizing Method:
Refer to this User Manual for details
EMC:
Group I Class A
Working System:
Continuous running equipment
Power Supply
Operating Voltage:
100V-240V AC
Line Frequency:
50/60 Hz
P
max
:
110VA
Fuse:
T1.6AL 250V
Environment
Monitor
Working
Temperature: 5 ºC ~ 40 ºC ( 41 ºF ~ 104 ºF)
Relative Humidity: 25% ~ 80% (non-condensing)
Atmospheric Pressure: 860hPa ~ 1060hPa
Transport and Storage
Temperature: -20 ºC ~ 55 ºC (-4ºF ~ 131 ºF)
Relative Humidity: 25% ~ 93% (non-condensing)
Atmospheric Pressure: 700hPa ~ 1060hPa