F9 Fetal & Maternal Monitor • User Manual
- 29 -
4.9 Patient Alarm Defaults
Alarm Setting
Options
Default
FHR1/FHR2 Alarm
On, Off
On
FHR1/FHR2 Lower Limit
50 ~ 205 bpm, adjustable in 5 bpm steps
120 bpm
FHR1/FHR2 Upper Limit
55 ~ 210 bpm, adjustable in 5 bpm steps
160 bpm
FHR1/FHR2 Alarm Delay
0 ~ 300 second(s), adjustable in 5 second steps 10 seconds
FHR1/FHR2 Alarm Level
Medium, not adjustable
Medium
NOTE: The upper limit must be higher than the lower limit. When setting the upper limit,
you do not have access to the options that are lower than the preset lower limit, and
vice versa.