F9 Fetal & Maternal Monitor • User Manual
- 27 -
4.3 Visual Alarm
When an alarm is active, the following 3 visuals occur:
-
Alarm indicator:
the alarm indicator flashes in orange with a frequency of 0.5 Hz if it is a
medium level alarm; the alarm indicator lights up continuously in orange if it is a low level alarm.
-
Alarm message:
the alarm message appears in the message window of the main interface in red,
with patient alarms on the left and technical alarms in the middle.
-
Flashing number:
the numerical value of the measurement flashes in grey with a frequency of 2
Hz.
When more than one alarm is active, the alarm messages appear in the same area in succession.
The patient alarm messages are displayed either:
•
in text form, for example “** FHR2 LOW”; or
•
in numeric form, for example “** FHR2 115 < 120”; ** indicates this is a medium level
alarm event; the first number is the current measurement result; the second number is the
preset alarm limit.
The technical alarm messages are displayed in text form, for example “Fetus EQUIP MALF”.
4.4 Choosing the Alarm Display Form
To change the patient alarm display form,
1) Select the Setup key
on the main interface.
2) Select
Alarm
Message
Form
.
3) Select
Text
(default) or
Numeric
.
4) Select
OK
.
4.5 Changing the Alarm Volume
To change the alarm volume,
1) Select the Setup key
on the main interface.
2) Select
Alarm
Alarm
Volume
.
3) Select
Low
(default),
Medium
or
High
.
4) Select
OK
.