16
Vacuum Pump Instructions for Use
8. Technical Data
Classification according to IEC 60601-1
Type of protection against electric shock:
Class I equipment
Degree of protection against electric shock:
Type B
Degree of protection against harmful ingress of solids and water:
IP41
Specifications
Power Supply:
100 - 240 VAC
Frequency:
50 - 60 Hz
Maximum current:
500 mA (115 VAC)
250 mA (240 VAC)
Maximum power consumption:
60 VA
Environmental operating conditions:
+5°C to +35°C
10% to 75% RH
700 hPa to 1060 hPa
Storage and transport directions:
+5°C to +40°C
10% to 75% RH
Manufactured and tested to the following standards:
IEC 60601-1: 1988 + A1: 1991 + A2: 1995
IEC 60601-1: 2005
IEC 60601-1-2: 2007
ISO 10079-1: 1999
VACUUM PUMP WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH:
ANSI/AAMI ES60601-1(2005 + C1: 09 + A2: 10),
CAN/CSA-C22.2 No. 60601-1(2008), E363262
Performance class:
High Vacuum / Low Flow (ISO 10079-1)
Dimensions:
200 mm wide x 100 mm high x 350 mm deep
Weight:
3.2 kg (7.1 lb)
Vacuum Ranges:
-10 mmHg to -500 mmHg in 1 mmHg increments.
-1.0 kPa to -67.0 kPa in 0.1 kPa increments.
Vacuum Range Accuracy:
±5 mmHg (±0.7 kPa)
Guidance and manufacturer’s declaration – electromagnetic emissions
The Vacuum Pump is intended for use in the electromagnetic environment specified below. The customer or the end user of the Vacuum Pump should assure
that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1
The Vacuum Pump uses RF energy only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Vacuum Pump is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies