18
Vacuum Pump Instructions for Use
9. Troubleshooting
IMPORTANT NOTE:
Should any errors persist, contact your COOK distributor.
Error and Alarm Indicator
Source of Error
Elimination of Error
Vacuum Pump will not turn on.
Displays do not illuminate.
Power supply cord is not connected. Mains
power switch is not turned on.
The device is in standby mode.
Check the power supply connection. (Refer to
§3.10).
Check that the Standby Indicator is illuminated
orange. Press the Standby Touch-Pad. (Refer to
§3.10).
Desired aspiration vacuum is not achieved.
Foot Pedal is defective.
Leak in the Vacuum Line or the Filter is wet.
Disconnect the Foot Pedal from the rear of the
device without applying pressure to the Foot Pedal
and re-connect the Foot Pedal.
Change the Vacuum Line and Filter. (Refer to §7).
Chime Volume Adjustment does not adjust.
Power supply cord is not connected. Mains
power is not turned on.
The device is not in standby.
Check the power supply connection. (Refer to
§3.10).
The device is functionally active, press the Standby
Touch-Pad to place the device in standby. (Refer
to §3.15).
Vacuum Pump displays the Vacuum in wrong
units.
The Display Units setting is set to either
mmHg or kPa.
Set the Display Units setting to desired units, either
mmHg or kPa. (Refer to §3.12).
Vacuum stays on when the Foot Pedal is
depressed and released.
The Foot Pedal Function is set to Latching.
Set the Foot Pedal Function to Non–Latching.
(Refer to §3.13.3).
Vacuum doesn’t stay on when Foot Pedal is
depressed and released.
The Foot Pedal Function is set to Non–
Latching.
Set the Foot Pedal Function to the Latching. (Refer
to §3.13.3).
Vacuum Pump displays an error code in the form
of ER1 to 5.
Device has an internal fault.
Contact service agent.
Recommended separation distance between portable and mobile RF communications equipment and the Vacuum Pump
The Vacuum Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the Vacuum Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Vacuum Pump as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people