background image

 

CO Sensor Calibration 
 

Required parts: 
 

 

 

Zero gas (20.9% O2) 

 

 

 

Span gas (250 PPM CO) 

 

 

 

3/16” I.D. rubber hose 

 

 

 

3LPM regulator 

 

 

 

UP99 Universal IR programmer 

These parts are included in the P1879-CO-250-NC-3 Calibration Kit. 

 
To calibrate the sensor the panel door must be opened so the user can view the LCD display on the 
Universal Transmitter and the adjustable flow regulator.  
 
The UP99 Universal Programmer is used to put the Universal Transmitter into Calibration mode.  To do so 
point the programmer at the infrared sensor on the top of the Universal Transmitter, then the unlock code 
must be entered.  To unlock the Universal Transmitter press 25 (the unlock code will not be displayed on 
the screen) then press enter (E) on the programmer.  Next enter the password which is factory defaulted to 
1 then press enter (E).  When the monitor is unlocked enter 111 then press enter to access calibration 
mode.  When in calibration the PLC will shut off the pump so it does not interfere with the calibration.  The 
Universal Transmitter interface is also explained in the Universal Transmitter Manual. 
 
Once in Calibration mode turn the handle of the ball valve on the right side of the enclosure to the upward 
position and connect the 3/16” I.D. hose from the 3LPM regulator on the “zero gas” cylinder to the 
calibration port on the right side of the enclosure.  The status light on the bottom of the Universal 
Transmitter will turn green when in calibration mode indicating that the zero gas should be applied to the 
sensor. Open the valve on the regulator and apply the zero gas. 
 
Make sure that the adjustable flow regulator inside the panel is fully open to allow maximum flow.  Apply 
the zero gas for two minutes.  After two minutes or a stabilized reading, point the IR programmer at the 
Universal Transmitter and press enter. The status light will now turn red indicating the monitor is ready to 
have the “span gas” applied.  Remove the zero gas and apply the span gas for two minutes.   
 
Notice the LPM level on the flow regulator; it should be approximately 3 LPM. The sample flow rate will 
have to be adjusted to match this flow for the most accurate sampling possible.   After two minutes or a 
stabilized reading, point the IR programmer at the Universal Transmitter and press enter.  The monitor will 
display if the calibration has passed or failed and go into standby mode for 60 seconds to allow the PPM 
level to drop to avoid false alarms. 
 
 Immediately after you have finished the span gas calibration, while the monitor displays either “Cal Pass” or 
“Cal Fail” quickly turn off and remove the span gas and turn the handle of the ball valve back to sample 
mode.  The PLC will automatically enter startup mode and turn the pump back on while the monitor is in 
standby mode to lower the gas value.  
 
At this point the adjustable flow regulator will need to be adjusted to match the 3LPM flow rate that the 
sensor was calibrated at. 
 
When the sample panel is in calibration mode the pump is turned off and all activities such as high pressure 
blow back and low flow blow back are disabled. The Universal Transmitter calibration is also explained in 
the Universal Transmitter Manual. 

Summary of Contents for CS0352-MP

Page 1: ...CONSPEC CONTROLS CS0352 MP User Manual 1 24 2017 Rev 3 ...

Page 2: ...urpose 1 System Overview 1 CS0352 MP Components 2 Panel View Screens 3 Modes 6 Trouble Conditions 7 CO Sensor Calibration 9 Flow Detector Calibration 10 See also 10 Replacement Parts 10 Contact Information 10 ...

Page 3: ...to detect smoldering coal that may have fallen to the bottom of the mill The pump is disabled and sampling ceases during a deluge or spray down of the mill to avoid the system from taking in too much water The PLC determines the mill status by several inputs from the MillPro system and acts accordingly The CS0352 MP performs scheduled high pressure blow backs to clear the sintered metal probes of ...

Page 4: ...Moxa Ethernet Switch RS485 to Serial Adaptor Probe Select Solenoid Pressure Release Solenoid H P Blow Back Solenoid Thermostat Trouble Relay Cooler Solenoid Relay 4 5 0LPM Flow Regulator Coalescing Filter Cooling Coil 3 way Ball Valve Calibration Calibration Port Exhaust Port 120VAC Supply Input System Inputs 4 20mA CO Level Output 1766 L32BWA SDR 120 24 900732 03 P2261 3 900724 AL 6SB P1907 SL SP...

Page 5: ... Cooler Solenoid Vortec Cooler Vortex 747 System 2711C T6T Vortex 747 System Vortex 747 System PanelView Screens Main screen The main screen contains a link to all the other screens All the screens are accessible from the main screen The screens are Maintenance Overview PV Config Alarm Log ...

Page 6: ... information all on one screen for easy viewing The panel status sensor status an approximated 4 20mA value CO value hour long trend window MillPro status Mill Status and logging information are all displayed The logs keep track of operation times so the user can predict when components will need maintenance or replacement The About screen can be accessed from the overview screen The about screen ...

Page 7: ...s such as a blocked probe during start up mode the PLC immediately goes into alarm and energizes the trouble relay This error is displayed on the panel view screen and needs to be cleared or acknowledged A record of alarms and errors is kept in this screen and can be cleared if desired ...

Page 8: ...t Up Mode Start up mode is the first mode that the sample panel goes into upon being enabled Start up mode allows pressure to be created within the lines and give the flow detector time to stabilize The sample panel enters start up mode every time that the pump turns on If there is a low flow condition or the line is clogged during start up mode the sample panel goes into an error condition and th...

Page 9: ...Mode Maintenance Mode should be entered by the user anytime that work is being performed on the system This will ensure that the scheduled blow back to occur every 30 minutes will not happen and the pumps and solenoids will be turned off unless forced on by the user Maintenance Mode is a useful tool for trouble shooting and checking for proper operation of individual components From this mode the ...

Page 10: ...n the user definable time frame Error will clear once panel is put into startup mode Alarm 1 The gas value has exceeded the alarm 1 user definable set point Alarm 2 The gas value has exceeded the alarm 2 user definable set point Temperature error The internal temperature of the panel is outside of the 4F 120F range of the gas sensor Flow Cal Error The pump is off but the flow detector still signal...

Page 11: ...d be applied to the sensor Open the valve on the regulator and apply the zero gas Make sure that the adjustable flow regulator inside the panel is fully open to allow maximum flow Apply the zero gas for two minutes After two minutes or a stabilized reading point the IR programmer at the Universal Transmitter and press enter The status light will now turn red indicating the monitor is ready to have...

Page 12: ...igher or lower low flow set point Once a successful calibration has been accepted lower the flow slowly to make sure the indicator light turns red at approximately 1 1 5 LPM and that you hear the relay energize and click If the calibration worked correctly raise the flow to approximately 3 LPM and close the door and put the panel back into run mode If the light does not turn red within 5 LPM of th...

Reviews: