PC-POD-CP-006-v01
PC-POD-CP-006-v01
Printed copies are not controlled.
Page
29
of
35
NOTE: When acquiring cardiac output data, bear in mind that the BP values are only
retrieved from the BP frame at the very end of the EA period.
9.3.17
Check the “Automatically save EA Data to Disk” to have acquired data saved.
9.3.18
Check the “Use PEP value from PEP Detector (B-wave cursor placement)” to
be able to compute systolic time interval values.
9.3.19
Click
to store and retain the options selected. These will be retained
through all subsequent executions of the software until changed by the user.
9.3.20
The “Data Acquisition Window” will appear.
9.4
OPTION 3 – Collect a single sample at a time
9.4.1
For acquisition of impedance data upon user command according to
specifications on the “Acquisition Setup Form” select “One-shot Mode” under
“Acquisition Mode”.
9.4.2
Under the “Ensemble Average MODE” select the mode by which data is to be
collected.
9.4.3
Under “Input-Digital-Filter-Setting”, select filter.
9.4.4
RRI options are used to calculate heart rate variability by measuring the
distances between the R peaks in the QRS complex. Under “RRI Data
Acquisition” select “RRI Not Active” to disable RRI data acquisition OR select
“Continuous RRI Mode” to enable RRI data acquisition during pre-selected
time intervals OR select “With Each EA Sample using HRV frequency…”.