9.3.5
Under “Input-Digital-Filter-Setting”, select filter.
9.3.6
RRI options are used to calculate heart rate variability by measuring the
distances between the R peaks in the QRS complex. Under “RRI Data
Acquisition” (RRI = R-R Interval) select “RRI Not Active” to disable RRI data
acquisition OR select “Continuous RRI Mode” to enable RRI data acquisition
during pre-selected time intervals OR select “With Each EA Sample using HRV
frequency…”. Note that the latter option can only be selected if the inter-
sample interval time is > 1.2 minutes.
9.3.7
If the “Continuous RRI Mode” is selected and the “No Limit” button appears,
acquisition will be continuous from the time the user starts data acquisition,
until the user stops the acquisition.
9.3.8
In order to set a time limit, click the “No Limit” button and a Dialogue Box
will appear.