54
NOTE:
To fine tune the reading to match the
reading on the o
xygen monitor, use the “Set 20mA
Span” menu. Push the joystick up or down until
both displays match.
CAUTION:
If the oxygen monitor is in the measuring mode and the ZERO push button is pressed
an “Err” will be displayed on the remote display and no readings will be recorded.
If the SPAN
push button is pressed in the measuring mode, the remote display will automatically adjust
to 25.0%.
You will need to repeat the above adjustment procedure again for the remote display to
mirror the oxygen monitor.
NEVER PUSH THE SPAN
in the measuring mode.
How to set the internal alarms on the Remote Display
The Remote Display has two user selectable alarms with a separate horn tone to identify
Alarm 1 from Alarm 2. Both are preset at the factory; AL1 = 19.5% and AL2 = 18%. They
can be changed in the field.
To set Alarm 1, Press Alarm 1 for two seconds to enter the alarm change mode. The
display will show the current alarm setting and the right most decimal point will be lit.
Use the UP or DOWN buttons to increase or decrease the alarm value. When done,
push the Alarm 1 button for 2 seconds to get back to the monitoring mode.