36
NONLATCH
O2 20.9%
LATCHING
O2 20.9%
..Set Latching-Relay 1, CO2…
O2 20.9%
..Set Latching-Relay 2, O2…
O2 20.9%
NONLATCH
O2 20.9%
LATCHING
O2 20.9%
..Set Latching-Relay 2, O2…
O2 20.9%
This is the menu at which to adjust the CO
2
alarm relay state on the dual O2/CO2 Monitor.
To change this value, push the joystick right to display the relay state. The display will
indicate
NONLATCH.
Pushing the joystick down will change the relay state from
NONLATCHING to LATCHING.
After entering the relay setting push the joystick left to enter the setting. The display will
scroll the following:
Push the joystick down to access the next sub menu;
Set Latching-Relay 2, O2,
will scroll
on the display. Use this menu to adjust the O
2
alarm relay state on the dual O2/CO2 monitor.
To change this value, push the joystick right to display the relay state. The display will
indicate
NONLATCH.
Pushing the joystick down will change the relay state from NONLATCH
to LATCHING.
After entering the relay setting push the joystick left to enter the setting. The display will scroll
the following: