9
ITEM
NAME
FUNCTION
1
VOLTAGE
Adjust Knob
Adjust the voltage set point in the tester.
Press it once to adjust the most significant digits and twice for the less significant digits, then turn Clockwise to increase the
output voltage before the
CHARGE
button has been pressed.
2
POLARITY
switch
Polarity is Fixed at positive. Having this button pressed while turning the tester On, the meter will display software version.
3
NOR REV
indicator
Indicates the state of the Output Polarity switch.
NOR
indicates Normal (Positive) position.
REV
indicates Reverse (Negative) position.
4
CHARGE
switch
Starts the charge process of the tester capacitor.
The
CHARGE
indicator will turn off after the
CHARGE
switch is pressed, and the
TRIGGER
indicator will turn on.
The charge process will stop after 2 minutes if the
TRIGGER
button is not pressed.
5
CHARGE
indicator
This Yellow indicator is lit to show that pressing the
CHARGE
switch is the next logical step in a test sequence.
CHARGE
indicator is lit when the tester is turn ON an after pressing
TRIGGER
button.
CHARGE
indicator will go out after pressing
CHARGE
button.
CHARGE
and
TRIGGER
Indicators will be blinking if the Interlock Switch is open. (Only testers with Interlock Switch
Option)
6
TRIGGER
switch
Triggers the output impulse waveform.
The impulse waveform will appear across the output leads.
7
TRIGGER
indicator
This Red indicator is lit to show that the tester can be trigger.
TRIGGER
indicator is lit for 2 minutes after the
CHARGE
button is pressed.
TRIGGER
indicator will go out after pressing
TRIGGER
button.
TRIGGER
and
CHARGE
Indicators will be blinking if the Interlock Switch is open (Only testers with Interlock Switch
Option)
TRIGGER
indicator will blink at when the Voltage. This effect will remain on until the
TRIGGER
switch is pressed.
(Only testers with PC Interface option)
8
VOLTAGE
meter
Displays the output voltage set point.
The voltage reading will increase from zero to the voltage set point when the CHARGE button is pressed. Note that the
Voltage meter may indicate that some residual voltage is present on the main storage capacitor, even when the tester is first
turned ON. This is due to inherent charging of the internal capacitors. Pressing the
TRIGGER
switch will discharge the
capacitors. Note that the peak amplitude of the measured output waveform is proportional to the voltage that is read of the
front panel of the MegaPulse, but it will always be somewhat lower. This is because the meter on the MegaPulse is
measuring the voltage on the main impulse storage capacitor. This voltage will intentionally dissipate to some extent before
reaching the output leads.
The meter will start to flash at
1600V
to indicate that voltage is in the maximum limits.