Safety Techniques 

 

The high voltage circuit of the MegaPulse Capacitor series A PF can be shut off at any time by turning 
OFF the rear power switch.  Always press TRIGGER to discharge the tester before turning OFF. 

 

The  MegaPulse  tester  is  provided  with  a  digital 

VOLTAGE  ADJUST

  knob  on  the  front  panel.    This 

voltage setting should always confirm by pressing the VOLTAGE ADJUST knob before start any testing.   

 

The MegaPulse tester is provided with a 

CHARGE

 switch that is in the unarmed "Standby" setting when 

the tester is first turned ON.  When the yellow 

CHARGE

 button is lit, the tester will not provide high 

voltage until the 

CHARGE

 Button and the 

TRIGGER

 Button have been pressed in order.  To prevent 

inadvertent operation, the operator should be instructed not to press the 

CHARGE

 Button until the test is 

ready. 
 
The  MegaPulse  tester  has  been  designed  for  one-touch  operation  with  the  right  hand.    If  possible,  it 
should be set up to the left and in front of the equipment under test.  The equipment under test should be 
connected to the MegaPulse tester and then left alone by the operator.  After the operator is clear of the 
Tester and the equipment under test, the operator should turn the rear-panel power switch to ON, confirm 
or adjust the Voltage Set Point, then press the 

CHARGE

 Button and wait until the front meter reaches 

the voltage selected, then press the 

TRIGGER

 Button, with his right hand.  This will allow the greatest 

separation between the operator and the test being conducted. 
 
The MegaPulse PF tester is designed to bleed the high voltage away after the test has concluded.  In order 
to  ensure  that  any  voltage  present  in  the  equipment  being  tested  has  been  completely  bled  away,  the 
operator  should  not  unplug  the  equipment  under  test  from  the  MegaPulse  until  the  front  panel  meter 
reads a safe level (40 volts or less is generally considered a safe level).  Pressing the TRIGGER button 
before disconnecting main power (or turning the equipment off) will ensure that the internal capacitors 
are discharged as much as possible. 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for MegaPulse Capacitor A PF Series

Page 1: ...MegaPulse IMPULSE TESTER Capacitor series A PF Instruction Manual...

Page 2: ...ign and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review thi...

Page 3: ......

Page 4: ...Megapulse Capacitor series A PF Specifications 7 Front and Rear Panel Features 8 Figure 1 Controls Indicators Connectors MegaPulse Capacitor series A PF Front Panel 8 Table 1 Controls Indicators Conn...

Page 5: ...1...

Page 6: ...re that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especial...

Page 7: ...d be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of...

Page 8: ...s rated for the voltage involved and that the frequency response of the instrument and probe are capable of measuring the output waveform of the MegaPulse Impulse Tester A measuring instrument or prob...

Page 9: ...F 1 User Manual 1 High Voltage Test Lead Red 00 HVL7R 3 feet 1 High Voltage Test Lead Black 00 HVL7B 3 feet 1 18 AWG AC Power Cord 70 101 1 TestMinder Optional User Manual 1 Software CD Megapulse Test...

Page 10: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Page 11: ...point adjusted by frontal VOLTAGE knob or by PC optional Polarity Control Positive only Voltage Display 4 Digit LED Display Voltage Meter resolution 5V Duty Cycle 1 pulse every 10 seconds Line Voltage...

Page 12: ...se of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 1 and described in Table 1 The rear panel features of the MegaPulse are shown in Figure 2 and...

Page 13: ...ross the output leads 7 TRIGGER indicator This Red indicator is lit to show that the tester can be trigger TRIGGER indicator is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator w...

Page 14: ...d to the chassis of the MegaPulse and is referenced to earth ground as long as the MegaPulse is properly grounded Even though this jack is referenced to ground it should be treated as hazardous whenev...

Page 15: ...d OFF 2 Fuse replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer inter...

Page 16: ...ect the ends of the test leads to a capacitor load of the range selected on the front panel knob plug your high voltage oscilloscope probe to the output leads as well Note that the RETURN lead is refe...

Page 17: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 18: ...ll be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or...

Page 19: ...ation procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Capacitor series A PF tester...

Reviews: