Unanticipated Results
The HTT-1R test may cause your tester to give results not consistent with those anticipated by
the switch setting of the HTT-1R. Until the problem is resolved, your tester should be removed
from service.
If there is an unexpected leakage or ground result, first check the hipot or ground bond trip limit
setting are correct.
If it is not detecting hipot failures, make sure testing voltage is above 1150Vac.
If there is a question regarding the proper operation of the HTT-1R, troubleshooting information
in Section 4 can be used to verify proper operation of the HTT-1R.