24
Section 5
Maintenance and Calibration
WARNING
THESE SERVICE INSTRUCTIONS ARE FOR USE BY QUALIFIED
PERSONNEL ONLY. TO AVOID ELECTRIC SHOCK, DO NOT
PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN THE
OPERATING INSTRUCTIONS UNLESS YOU ARE QUALIFIED TO DO
SO.
Introduction
This section of the manual contains maintenance information for the HT-2000P Dielectric
Withstand Tester. This maintenance information is divided into service information, general
maintenance, a performance test, and a calibration procedure. The performance test is
recommended as an acceptance test when the instrument is first received, and later as a
preventative maintenance tool to verify proper instrument operation. A 1-year calibration cycle is
recommended to maintain the specifications given in Section 1. A three inch piece of approx. 18
AWG insulated hookup wire with each end stripped back ½" and a lead with a banana plug on
one end are required to conduct the performance test. The test equipment required for the
calibration procedure is a DMM able to read true rms 0-2000 Vac
±
1% and a 3-ohm 1/4 w 1%
resistor.
Service Information
The HT-2000P is warranted to the original purchaser for a period of 1 year. This
warranty does not cover problems due to misuse or neglect.
Malfunctions which occur within the limits of the warranty will be corrected at no
charge. Mail the instrument post paid to the manufacturer. Dated proof of purchase is required
for all in-warranty repairs.
The manufacturer is also available for calibration and/or repair of instruments that are
beyond their warranty period. Contact the manufacturer for a cost quotation. Ship the instrument
and your remittance according to the instructions given by the manufacturer.
General Maintenance
Interior Access
NOTE
To avoid contaminating the PWB with oil from your fingers, handle it by the edges or
wear gloves. If the PWB becomes contaminated, refer to the cleaning procedures given
later in this section.