background image

IE-LOG-S5141-01 

 

  3 

Total memory capacity:   
  for non-cyclic mode  

8 126 records for each measured channel 

for cyclic mode 

 

7 880 records for each measured channel 

  Specified values are maximum possible (when all channels are switched on) and can be 

reached only if record is not interrupted (since last memory erasing)  

Communication with computer: via RS232 (serial port) by means of COM adapter or USB 

port  by  means  of  USB  adapter;  data  transfer  from  logger  via  communication  adapter  is 
optical 

Real time clock: adjustable from computer, integrated calendar including leap years 
Error of internal RTC: < 200 ppm (i.e. 0.02 %, 17.28 s in 24 h) 
Power: Lithium battery 3.6 V size AA 
Typical battery life: 

typical (data download to PC is carried out about once a week): 5 years 
in continuous on-line mode with interval 1 minute: approximately 3 years 
in continuous on-line mode with interval 10 seconds: approximately 1 year 
Notice:  The  above  battery  lives  are  valid  for  logger  operation  in  temperature  from  -5  to 
+35°C. If often operated outside of the above temperature range battery life can be reduced 
to 75%. 

Protection: IP20 
Operational conditions: 
   Operational temperature range: -30 to +70 °C 
  Operational humidity range: 5% to 85 %RH 
   Specification  of  outer  characteristics  accordingly  to  Czech  National  Standard  33-2000-3: 

normal environment accordingly to appendix NM:  AE1, AN1, AR1, BE1 

   Operational position: negligible  
  Logger installation: by self adhesive Dual Lock, applied to clean, flat surface 

Not  allowed  manipulation:  incoherent  surface  is  unsuitable  for  installation  of  the  logger 

using the self adhesive Dual Lock 

Limit condition:  temperature -40 to +70 °C, humidity 5 to 85 %RH 
 

maximum voltage at measuring input (IN1, IN2 and IN3) -1 to +7V dc 

Storing condition: temperature -40 to +85 °C, humidity 5 to 90 %RH 
Dimensions: 93 x 64 x 29 mm 
Weight including battery: approximately 130 g 
Material of the case: ABS 

 

 

Logger operation 

Logger comes complete with installed battery and switched OFF. Before operation it is necessary 

by means of installed user PC software to set logging parameters and other features. The PC program 
is common for all types of loggers Sxxxx – 

for type S5141 it is necessary to have installed version 

4.0.17.0

  or  higher.  Older  program  version  can  cause  logger  reconfiguration  together  with  loss  of 

proper  function!  Installed  program  version  can  be  verified  in menu  Help

  /  About  program…

  Free 

newest program version is available to download at 

www.cometsystem.com

. For communication with 

the PC a communication adapter is necessary (not included in delivery). For connection via RS232 
serial port it is necessary to use COM ADAPTER, for connection via USB port it is necessary to use 
USB  ADAPTER.  Connect  adapter  connector  to  proper  computer  port  and  plug  the  adapter  to  the 
guide slots on the side of logger.  

 
 
 
 

Summary of Contents for S5141

Page 1: ...LOGGER S5141 Three channel 0 5 Vdc datalogger with internal temperature sensor and display Instruction Manual...

Page 2: ...nction Also memory alarm mode is selectable when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each channel individually Logging mode can be adjusted...

Page 3: ...normal environment accordingly to appendix NM AE1 AN1 AR1 BE1 Operational position negligible Logger installation by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation inco...

Page 4: ...er automatic switching ON or disable this option select if record runs permanently or only if alarm is active if alarms are about to be applied set both limits for each measured quantity and enable al...

Page 5: ...the display in usual operation logger switched ON After switching ON the logger all LCD symbols are displayed for checking the display Then actual date and time in logger is displayed for approximate...

Page 6: ...bol is replaced by adjacent symbol hyphen It appears in case all measured values are inside of adjusted alarm limits and data logging therefore does not run Displayed symbol indicates logger is ON If...

Page 7: ...lled in non cyclic logging mode logger is switched OFF and message MEMO FULL appears on the LCD It appears also if logger is operated with switched OFF display New initialization of logger can occur i...

Page 8: ...ogger OFF by magnet The procedure is identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Reset of...

Page 9: ...cially the real time clock in logger Attention inserted battery with incorrect polarity causes logger damage put the lid back again and screw four screws Be careful to have the rubber sealing properly...

Reviews: