INIT-Initialize Memory
The INIT menu allows the operator to selectively clear data stored for
program, calibration and/or clock functions. To select one of the
functions, simply press the SET key when the upper display shows the
INIT menu.
Next, use the UP/Down arrow keys to select which data to clear.
The choices are as follows:
1. VAR Miscellaneous Variables
2. PRG Program Data
3. TIM Time Data
4. CBR Calibration Data
Finally, with the desired data type displayed in the lower display, press the
SET key again to reset the data to the factory set values. The lower display
will show CLR indicating the data has been cleared.
6-12
Vacuum Oven
Cole-Parmer
Section 6
Operation
Table 8: Calibration Menus
Prompt
Calibration Routine
Comments
INIT
Initialize data memory.
Four options. Selectively clears data
variables, program and clock. Resets all
calibration data to defaults.
TCCB
Measure and Control thermocouple
calibration.
Requires two temperature points (T2 -T1)
>100°C temperature control loop is on
during calibration.
VCMC
Pressure sensor.
Requires two pressures, ambient plus
second point near 0 in Hg.
D/AC
Temperature and vacuum recorder
outputs
Adjust full scale. Temperature, vacuum
inputs not required.
CJCD
Cold junction compensation temp.
Display for verification only. No adjust-
ment