Follow the instructions below to access the calibration mode:
1. Set the POWER switch to the OFF position.
2. Remove the right side panel.
Note
Refer to instructions for removing the side panel in Auxiliary Pump
Contacts before proceeding.
s
3. Locate two sets of jumpers on the back of the control board labeled
JP4. To enter the calibration mode, remove the bottom jumper and
place it across the bottom pins as shown in the control board
illustration below.
4. Set the power switch to the ON position. The upper display will show
INIT, which is the first calibration menu. By pressing the MENU key
repeatedly, all calibration menus can be viewed in the upper display.
Table 8 summarizes the calibration menus.
Note
Any part of the calibration can be conducted individually without
altering data from the others.
s
5. To abort a calibration and return to the control mode, simply set the
power switch to the OFF position and remove JP4 before completing
a calibration. The previous calibration data will not be affected.
Vacuum Oven
6-11
Cole-Parmer
Section 6
Operation
Calibration Mode
Battery module
Calibration Jumper P4
(shown in Calibration mode)
Recorder
outputs - P3
Summary of Contents for 05017-10
Page 1: ...Stable Temp Model 282A Vacuum Oven 105714 7006281 Rev 4...
Page 8: ......
Page 12: ......
Page 36: ......
Page 40: ......
Page 44: ...Vacuum Oven 9 1 Cole Parmer Section 9 Electrical Schematic 115V Schematic 230V Schematic...