Troubleshooting

BioXp™ deck inspection error

After you load the deck and close the instrument door for 

a system run, the instrument performs a deck inspection, 

checking the location and orientation of the recovery plate, 

ethanol reservoir, reagent strip handles, reagent strips, tips, 

and system robotics.

Inspecting system

Inspecting recovery plate

Inspecting reservoir

Inspecting strip handles

If any anomalies are detected, the instrument displays a deck 

inspection failure message, similar to the screen shown on the 

right. If the system displays a deck inspection failure message, 

see the following table for recommendations.

Deck inspection found problems:

Plate found in purification station.

Recovery plate not found or is rotated.

Reservoir missing or misplaced.

Strip handle not closed.

Attention

Ignore and continue

Stop and fix

Issue

Possible solution and recommendations

Deck inspection detected anomalies with one (or more) of 

the following:
• Plate found in purification station
• Recovery plate not found or is rotated
• Ethanol reservoir is missing or misplaced
• Strip handle not closed

Press 

Stop and fix

• Open the door and correct the deck inspection issue.
• After fixing the error, close the door to initiate a second 

deck inspection.

• If the deck inspection error message persists after you 

have corrected any issues, call Codex DNA technical 

services.

Error messages

If you observe an error screen during operation, follow the 

screen instructions. If directed or if the problem persists, make 

a note of the screen and call Codex DNA technical services. 

Additional information is included in the following table.

 

Read job issue. Open the door to retry.

If this persists, call 855.474.4362.

Attention

Unlock door

General troubleshooting

Codex DNA technical services: 858.228.4115

Issue

Possible solution and recommendations

Touchscreen does not light up

• Confirm that the power switch is in the 

ON

 position.

• Confirm that the instrument is plugged into a working 

electrical outlet.

• Verify that the plug is securely connected to the side of 

the instrument.

• Contact Codex DNA technical services.

Power failure occurs during the run

Contact Codex DNA technical services.

Difficulty operating the touchscreen

Cleaning and maintenance

Cleaning instructions

• Clean the exterior of the BioXp™ system with standard cleaners and a damp cloth.
• Clean the internal surfaces periodically using a mild detergent or 70% ethanol. Spray the cleaner onto a cloth and then use the 

damp cloth to wipe instrument surfaces. Do not spray any cleaners directly into the instrument.

Do not use abrasive or caustic cleaners on internal or external instrument surfaces.

Spills

Take every precaution to ensure that no liquids are spilled in the instrument. Wipe up accidental spills in and around the 

instrument immediately. If ethanol spills, dispose of ethanol-saturated towels according to your institutional guidelines.

In the event of hazardous material spillage

• In the event of spillage of a hazardous material, the BioXp™ system user is responsible for decontamination of the hazardous 

material on or inside the instrument.

• Before using a cleaning or decontamination method other than a recommended cleaner, contact Codex DNA to verify that the 

proposed decontamination method will not damage the equipment.

Summary of Contents for 43029

Page 1: ...1 BioXp 3250 system User guide Part number 43029 Revision 1 1 Effective 11DEC2020...

Page 2: ...IONS EXTENSIONS OR ADDITIONS TO THIS WARRANTY e You shall indemnify defend and hold harmless Codex DNA Inc from any costs expenses damages or other losses arising out of i any warranty of greater scop...

Page 3: ...standards Alternating current Waste Electrical and Electronic Equipment WEEE directive Do not dispose symbol For more information see Equipment disposal and WEEE compliance Consult instructions for u...

Page 4: ...d to an extension cord or to a multiple portable socket Doing so may compromise shielding and or grounding This equipment should not be used adjacent to or stacked with other equipment If it becomes n...

Page 5: ...tem communication 3 Fuse Do not use 4 Power switch Switch power to the instrument on or off Note Leave the instrument on continuously 5 Power inlet Power the instrument BioXp applications The BioXp 32...

Page 6: ...ent is unable to perform a successful self check it will alert the user to the problem and will prevent the instrument from operating until the problem has been addressed The self check process has be...

Page 7: ...s 2 Add a minimum of 12 mL freshly prepared 70 ethanol to the reusable ethanol reservoir a Load ethanol reservoir in the right most reservoir retainer position of the instrument deck Note Do not disca...

Page 8: ...ep running to continue the run Are you certain you want to cancel this job You will not be able to restart it Attention Yes Cancel job No Keep running If you press YES Cancel job the instrument will d...

Page 9: ...technical services Additional information is included in the following table Read job issue Open the door to retry If this persists call 855 474 4362 Attention Unlock door General troubleshooting Code...

Page 10: ...instructions may cause harmful interference to radio or television reception However there is no guarantee that interference will not occur in a particular installation If this equipment does cause in...

Page 11: ...power quality should be that of a typical commercial or laboratory environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for...

Reviews: