Safety information
Manual symbols
This manual describes the use and maintenance of the Codex DNA BioXp™ 3250 system. Read this manual completely before
putting the instrument into service. The following icons are used in this manual:
Electrical safety warning:
Failure to follow these instructions can cause harm to operators or subjects.
Important note:
Failure to follow these instructions can lead to damage of the system or unexpected results.
Precautions
The BioXp™ system conforms to the following standards:
• IEC 61010-1:2010 third edition, EN 61010-1:2010, UL, CSA
• EMC: IEC 61326-1:2012; EN 61326-1:2013
• KN 61000-6-4: 2015 and KN 6100-6-2: 2015, AUS/NZ CISPR 11
The radiation exposure hazard is non-critical.
Covers and exterior components which the operator may come in contact with during routine maintenance or
calibration, shall operate at a voltage no greater than 25 V AC or 60 V DC.
Marking symbols on the equipment
Warning:
Indicates a warning concerning operations that may lead to personal injury or
potential damage to the instrument if not performed correctly.
DO NOT MOVE
BY HAND
CAUTION
Caution, hot surface, do NOT touch.
Conformité European:
This mark is a declaration by the manufacturer that the respective
component complies with the relevant directives and standards as issued by the European
Union.
Korean EMC compliance:
This symbol indicates that the product conforms to relevant South Korean
EMC standards.
사용자안내문
사용자안내문
이기기는업무용환경에서사용할목적으로적합성평가를받은기기로서가정용환경에서사용하는
경우전파간섭의우려가있습니다.
사용자안내문은 “업무용방송통신기자재”에만적용한다.
Alternating current.
Waste Electrical and Electronic Equipment (WEEE) directive:
Do not dispose
symbol. For more
information, see
Equipment disposal and WEEE compliance
Consult instructions for use.
Warning:
Flammable material.
Safety warnings
Installation
• The BioXp™ system is heavy. Do not attempt to lift or move the instrument without the assistance and
supervision of colleagues. Be certain to use appropriate moving equipment and proper lifting techniques
(including but not limited to making certain that you have a secure grip on the instrument before lifting,
keeping your spine in a neutral position while lifting with your legs, and coordinating lifting and moving
movements with all appropriate personnel). Improper lifting can cause permanent back injury.
• Use this product only as specified. If the instrument is used in a manner not specified by Codex DNA,
the protection provided by the instrument may be impaired, resulting in personal injury or damage to the
instrument.
Ground circuit protection
• Check the mains electricity supply plug, voltage, and frequency to confirm that the supply corresponds to
the values shown on the instrument label.
• Ensure that the supply of power to the instrument is via a socket fitted with a protection device, such as a
circuit breaker of the correct rating, to provide automatic power cut-off in the event of an insulation fault.
• Confirm that the mains wiring is grounded. Codex DNA declines all responsibility for any damages due to
instrument non-grounding.
Main power switch
Do not position the instrument such that it is difficult to operate the main power switch.