Setting up the BioXp™ system
Leave the BioXp™ system powered on
We recommend leaving the instrument on continuously. Only turn the instrument off when necessary, such as when moving the
instrument to another location. If the instrument
must
be turned off, toggle the power switch to the
Off
position.
Ethernet setup
The BioXp™ system requires an internet connection. Prior to connecting the instrument to the internet, be sure to test the
connection. Connect the provided RJ45 ethernet cable to the ethernet port — located on the side of the unit — to an available
ethernet wall jack.
In order for the BioXp™ system to communicate with the Codex DNA server, you must open the 80/443 port through
your network. In addition, the BioXp™ system will be making an outbound SSL connection (HTTPS/SSL) to logmein.com,
codexdna.com and drive.google.com. Unblock or whitelist these sites.
Start the BioXp™ system
1. Ensure that all packaging materials have been removed, including the
shipping screw located under the pipette head.
2. Turn the instrument on using the power switch located on the side of the
instrument.
3. Close the instrument door. The BioXp™ system will initiate a self-check
process to verify the correct functionality of motion control and other
internal systems of the instrument. If the instrument is unable to perform a
successful self-check it will alert the user to the problem and will prevent
the instrument from operating until the problem has been addressed. The
self-check process has been carefully developed to ensure that the BioXp™
system is operating within optimal calibration ranges and that all systems are
functioning correctly before initiating a run.
Configure the BioXp™ system
1. Press
Configuration
Insert new job materials and
close door to start the next job.
Or change the configuration.
BioXp™ system ready
Configuration
2. Press
Network
Configuration
Date and time
Service
Network
Language
Done
Test the network connection
Press
Test connection
If the system displays
Network connection working normally
,
the BioXp™ system is ready for use. If you are ready to load
the instrument and begin a run, refer to the appropriate deck
loading instructions for your system run.
Network
Cancel
IP address
10.32.36.110
Subnet mask
10.32.36.110
Default gateway
10.32.36.110
Preferred DNS server
MAC address
10.32.36.110
MAC address
Use DHCP
Network connection
working normally
Test connection
OK
11:03:1D:12:8AF7
If the system displays
No network detected
, contact
Codex DNA customer service at
for
assistance in resolving the connection.
Network
Cancel
IP address
10.32.36.110
Subnet mask
10.32.36.110
Default gateway
10.32.36.110
Preferred DNS server
MAC address
10.32.36.110
MAC address
Use DHCP
No network detected
Test connection
OK