CODEX DNA 43029 User Manual Download Page 4

6

7

Hazardous chemical warning

• Chemicals used with the instrument may be hazardous.
• Always wear protective gear including safety goggles, gloves, and a lab coat when handling chemicals.
• Exercise caution when handling flammable liquids.
• Take precautions, including but not limited to installing proper ventilation systems, to ensure that the BioXp™ 

system laboratory is safe and that operators of the instrument will not be exposed to hazardous levels of 

harmful chemicals.

• Follow all national, state, and local health and safety regulations and laws.
• Use proper waste disposal in accordance with all relevant regulations.
• Refer to applicable material safety data sheets (MSDSs).

Flammable liquids: Instrument warning

• Exercise caution when using flammable liquids in the BioXp™ system to minimize the risk of fire.
• Do not use any flammable liquids in the instrument except for the liquids required in the included procedure.
• Provide adequate ventilation for the instrument to prevent the accumulation of flammable vapors.
• All sources of ignition, such as open flame or electrical spark, are not permitted near the instrument or in 

areas where flammable vapors may travel.

• Ensure that the instrument, all neighboring equipment, and users of the instrument are appropriately 

grounded to minimize the generation of static electricity.

• Do not use the ethanol reservoir if it is cracked.

Flammable liquids: Solution handling warning

• Keep flammable liquids in covered containers when not in use.
• Provide means to promptly and safely dispose of any flammable liquid leaks or spills.
• Do not transfer liquids using air pressure.
• Read applicable MSDSs.
• Store flammable liquids in cool, well-ventilated areas away from corrosives, oxidizers, and ignition sources.
• Label containers and cabinets as “flammable materials” where applicable.
• Use only approved safety vessels for flammable liquid storage.
• Ground and bond flammable liquid containers to prevent static charge buildup.
• Never pour flammable liquids down a drain or sink.
• Dispose of empty flammable containers in an approved manner.

Electrical safety warning

• Do not attach the power cord to an extension cord or to a multiple portable socket. Doing so may 

compromise shielding and/or grounding.

• This equipment should not be used adjacent to or stacked with other equipment. If it becomes necessary 

to use the BioXp™ system under these conditions, the unit should be observed to verify normal operation in 

this condition.

• Use only grounded mains outlets to supply instrument.
• Do not remove any panels. Panels should be removed only by trained service personnel.
• The AC power cord is the primary disconnection device for the instrument. Ensure that the point of 

connection is easily accessible and free of any obstructions.

Important warning: Moving parts

• Do not override the door lock. When the door is locked, the moving parts of the instrument are operational. 

Never attempt to physically restrict movement of instrument components.

• Interfering with moving parts may cause personal injury. Keep hands clear of moving parts while operating 

the instrument.

• Do not move or shift the instrument during operation.
• If the instrument shows evidence of corrosion or wear, do not attempt to manipulate or tamper with 

components of the instrument.

Important warning: Disconnecting the instrument

• Disconnect power to the system before cleaning or performing maintenance.
• DO NOT remove any panels; they should be removed only by qualified service personnel.
• Do not disassemble unit.

Equipment specifications

System components

• Use of parts or materials other than those supplied with the instrument can degrade system performance 

and may cause safety hazards. Use only Codex DNA approved or supplied components.

• Use only Codex DNA supplied power cord. Use of other power cords may compromise electrical protection 

and could create a hazard.

Components of the BioXp™ system

• BioXp™ 3250 system
• Power cord
• RJ45 ethernet cable
• Reagent and recovery plate thermal covers
• Ethanol reservoir

Technical specifications

Product name

BioXp™ 3250 system

Power input voltage

100–240 V

Power input current

8.3 A max

Operating temperature range

16 °C to 40 °C

Storage temperature range

−18 °C to 60 °C

Operating and storage humidity range

10% to 90% (non-condensing relative humidity)

Operating altitude

Up to 2,000 m

Water ingress

Non-immersion, protection for damp wipe only

Weight

63.4 kg [139.8 US lbs]

Dimensions (W × H × D)

69 × 77 × 53 cm [27 × 30 × 21 in]

Summary of Contents for 43029

Page 1: ...1 BioXp 3250 system User guide Part number 43029 Revision 1 1 Effective 11DEC2020...

Page 2: ...IONS EXTENSIONS OR ADDITIONS TO THIS WARRANTY e You shall indemnify defend and hold harmless Codex DNA Inc from any costs expenses damages or other losses arising out of i any warranty of greater scop...

Page 3: ...standards Alternating current Waste Electrical and Electronic Equipment WEEE directive Do not dispose symbol For more information see Equipment disposal and WEEE compliance Consult instructions for u...

Page 4: ...d to an extension cord or to a multiple portable socket Doing so may compromise shielding and or grounding This equipment should not be used adjacent to or stacked with other equipment If it becomes n...

Page 5: ...tem communication 3 Fuse Do not use 4 Power switch Switch power to the instrument on or off Note Leave the instrument on continuously 5 Power inlet Power the instrument BioXp applications The BioXp 32...

Page 6: ...ent is unable to perform a successful self check it will alert the user to the problem and will prevent the instrument from operating until the problem has been addressed The self check process has be...

Page 7: ...s 2 Add a minimum of 12 mL freshly prepared 70 ethanol to the reusable ethanol reservoir a Load ethanol reservoir in the right most reservoir retainer position of the instrument deck Note Do not disca...

Page 8: ...ep running to continue the run Are you certain you want to cancel this job You will not be able to restart it Attention Yes Cancel job No Keep running If you press YES Cancel job the instrument will d...

Page 9: ...technical services Additional information is included in the following table Read job issue Open the door to retry If this persists call 855 474 4362 Attention Unlock door General troubleshooting Code...

Page 10: ...instructions may cause harmful interference to radio or television reception However there is no guarantee that interference will not occur in a particular installation If this equipment does cause in...

Page 11: ...power quality should be that of a typical commercial or laboratory environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for...

Reviews: