14
15
Initiating a BioXp™ run
1. Once the deck is loaded, confirm that all components are
securely seated.
2. Close the door.
3. Press
Start
. The run will initiate.
Cancel
START
2-FRAGMENT CLONE 16S + AMPLIFY V1.1 EM
2-FRAGMENT CLONE 16S + AMPLIFY V1.1 EM
Tracking progress during a run
Time to completion
During a run, the instrument will display the projected time
remaining for the run (in hours : minutes : seconds) and the
estimated time of completion.
2-fragment clone + amplify V1.1 M
Time remaining
00:00:29
Time of completion
3:07:59
Slide to cancel
Step 1
Step 2
Step 3
Step 4
Cancelling a job during a run
IMPORTANT!
You will not be able to reuse reagents if you cancel a job after the run has initiated and the reagent plate seal has
been broken. Do not cancel a run unless absolutely necessary.
1. If you need to cancel a job during a run, slide your finger
to the right across the bar on the bottom of the display
screen during a run.
Slide to cancel
2. After sliding to cancel, the system will display a
confirmation to cancel screen. Press
YES — Cancel job
to
end the run or
No — Keep running
to continue the run.
Are you certain you want to cancel this
job? You will not be able to restart it.
Attention
Yes – Cancel job
No – Keep running
If you press
YES — Cancel job
, the instrument will display a
Canceling job
screen until the job cancel is complete.
Attention
Cancelling job
Storing covers
Opening thermal cycler
3. When prompted, remove all materials and used or
partially used consumables from the instrument deck.
Attention
Open door and remove materials
from a previous job (if any remain).
Then close door to continue.
BioXp™ run completion
1. When the BioXp™ run is complete, open the door.
2. All DNA products are now located in the recovery plate. The plate map provided with the project identifies specific locations
of each fragment and final product. Seal and store the recovery plate at 2 °C to 8 °C for up to one week or at −20 °C for up to
one year.
3. Empty and dry the ethanol reservoir.
4. Empty the tip waste and discard remaining consumables.
BioXp™ gene fragments and variant libraries
BioXp™ cloning projects
. . . .
BioXp™ 3200
Oligo vault
BioX
p™ DNA
assembly plate
P/N: BX1230
-XX
Lot: 123
MFG date: 09/01/2020
OLIGO VAULT™ PLATE
RECOVERY PLATE
BioXp™ 3200
Oligo vault
. . . .
BioX
p™ DNA
assembly plate
P/N: BX1230
-XX
Lot: 123
MFG date: 09/01/2020
OLIGO VAULT™ PLATE